FDA Enforcement Class I Ongoing

Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. MTS RH LFT HRT KIT, Medline SKU # 60010508; 2. MTS 3V W/GUIDE, Medline SKU # 60100701; 3. ANGIOGRAPHY KIT, Medline SKU # 60131209; 4. MTS LHK, Medline SKU # 60160075; 5. MTS LHK, Medline SKU # 60161954; 6. MTO LHK, Medline SKU # 60230148; 7. GENERAL SURGERY TRAY, Medline SKU # 00-500510I; 8. BIOBURDEN TEST KIT - 686, Medline SKU # BIOB686; 9. RADIO ANGIO UNIV HOSP PACK-LF, Medline SKU # DYNJ0376145M; 10. SPECIALS PACK-LF, Medline SKU # DYNJ0392706K; 11. CARDIAC CATH PACK-LF, Medline SKU # DYNJ0451426O; 12. CARDIAC CATH PACK-LF, Medline SKU # DYNJ0451426P; 13. CARDIAC CATH PACK, Medline SKU # DYNJ07953N; 14. CATH PACK-LF, Medline SKU # DYNJ0803367J; 15. E.P. TRAY, Medline SKU # DYNJ19343J; 16. CARDIAC CATH PACK, Medline SKU # DYNJ20898T; 17. CATH LAB TRAY, Medline SKU # DYNJ24250G; 18. CATH LAB PACK-LF, Medline SKU # DYNJ38333D; 19. CATH LAB PACK-LF, Medline SKU # DYNJ38333F; 20. EP IMPLANT PACK-LF, Medline SKU # DYNJ38910D; 21. ADULT CATH LAB PACK, Medline SKU # DYNJ40380J; 22. CRANIOTOMY PACK-LF, Medline SKU # DYNJ43403D; 23. LAP CHOLE PACK, Medline SKU # DYNJ48487F; 24. CATH LAB ANGIOGRAPHY PACK, Medline SKU # DYNJ50779G; 25. CATH LAB ANGIOGRAPHY PACK, Medline SKU # DYNJ50779I; 26. HYBRID, Medline SKU # DYNJ57049N; 27. SPECIALS PK, Medline SKU # DYNJ57713C; 28. SPECIALS PK, Medline SKU # DYNJ57713D; 29. SPECIALS PK, Medline SKU # DYNJ57713F; 30. CARDIAC CATH LAB, Medline SKU # DYNJ57748A; 31. TAVR PACK, Medline SKU # DYNJ58771G; 32. CATH PACK GWV, Medline SKU # DYNJ59229F; 33. HEART CATH PACK, Medline SKU # DYNJ61294F; 34. CATH LAB PACK, Medline SKU # DYNJ61353C; 35. DEVICE IMPLANT PACK, Medline SKU # DYNJ64636A; 36. CUSTOM PROCEDURE KIT, Medline SKU # DYNJ65277A; 37. GENERAL VASCULAR PACK RFID-V2, Medline SKU # DYNJ66442F; 38. CARDIAC CATH PACK, Medline SKU # DYNJ69845C; 39. OMF MINOR PACK, Medline SKU # DYNJ83991C; 40. HEART CATH PACK, Medline SKU # DYNJ84133B; 41. SPECIAL PROCEDURES, Medline SKU # DYNJ85425B; 42. PRIMARY PACK, Medline SKU # DYNJT3093; 43. CARDIAC CATH PACK, Medline SKU # DYNJT3522A; 44. CARDIAC CATH PACK, Medline SKU # DYNJT3522B; 45. TOTAL KNEE PACK, Medline SKU # DYNJT4069; 46. CARDIAC CATH PACK, Medline SKU # DYNJT5685; 47. CATH PACK, Medline SKU # DYNJT6544; 48. PACK CARDIAC CATH, Medline SKU # DYNJT7591; 49. OBL CATH VEIN PACK, Medline SKU # DYNJT7863; 50. TUMESCENT SYRINGE KIT, Medline SKU # DYNJTUMSYR; 51. CSTM LH KIT, Medline SKU # VASC1054; 52. LHK, Medline SKU # VASC1130A; 53. MANIFOLD KIT, Medline SKU # VASC1137A; 54. LHK, Medline SKU # VASC1299A; 55. LHK, Medline SKU # VASC1399B; 56. LHK, Medline SKU # VASC1403; 57. IR 2 PORT, Medline SKU # VASC1404; 58. LHK, Medline SKU # VASC1405; 59. LHK, Medline SKU # VASC1455A; 60. TWO PORT MANIFOLD KIT, Medline SKU # VASC1496; 61. RHK, US, NAMIC MANIFOLD, Medline SKU # VASC1502; 62. LHK, Medline SKU # VASC1555; 63. 4 VALVE,PERC,MANIFOLD,KIT, Medline SKU # VASC1609; 64. LHK, Medline SKU # VASC1668; 65. LHK, Medline SKU # VASC1694; 66. LHK, Medline SKU # VASC1805; 67. TVR, Medline SKU # VASC1816.

Recall: Z-1724-2026 · Reported April 15, 2026

Enforcement

Recall Number
Z-1724-2026
Event ID
98598
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Medline Industries, LP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
April 15, 2026
Initiation Date
February 27, 2026
Classification Date
April 7, 2026
Address
3 Lakes Dr, Northfield, IL, 60093-2753, United States

Description

Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. MTS RH LFT HRT KIT, Medline SKU # 60010508; 2. MTS 3V W/GUIDE, Medline SKU # 60100701; 3. ANGIOGRAPHY KIT, Medline SKU # 60131209; 4. MTS LHK, Medline SKU # 60160075; 5. MTS LHK, Medline SKU # 60161954; 6. MTO LHK, Medline SKU # 60230148; 7. GENERAL SURGERY TRAY, Medline SKU # 00-500510I; 8. BIOBURDEN TEST KIT - 686, Medline SKU # BIOB686; 9. RADIO ANGIO UNIV HOSP PACK-LF, Medline SKU # DYNJ0376145M; 10. SPECIALS PACK-LF, Medline SKU # DYNJ0392706K; 11. CARDIAC CATH PACK-LF, Medline SKU # DYNJ0451426O; 12. CARDIAC CATH PACK-LF, Medline SKU # DYNJ0451426P; 13. CARDIAC CATH PACK, Medline SKU # DYNJ07953N; 14. CATH PACK-LF, Medline SKU # DYNJ0803367J; 15. E.P. TRAY, Medline SKU # DYNJ19343J; 16. CARDIAC CATH PACK, Medline SKU # DYNJ20898T; 17. CATH LAB TRAY, Medline SKU # DYNJ24250G; 18. CATH LAB PACK-LF, Medline SKU # DYNJ38333D; 19. CATH LAB PACK-LF, Medline SKU # DYNJ38333F; 20. EP IMPLANT PACK-LF, Medline SKU # DYNJ38910D; 21. ADULT CATH LAB PACK, Medline SKU # DYNJ40380J; 22. CRANIOTOMY PACK-LF, Medline SKU # DYNJ43403D; 23. LAP CHOLE PACK, Medline SKU # DYNJ48487F; 24. CATH LAB ANGIOGRAPHY PACK, Medline SKU # DYNJ50779G; 25. CATH LAB ANGIOGRAPHY PACK, Medline SKU # DYNJ50779I; 26. HYBRID, Medline SKU # DYNJ57049N; 27. SPECIALS PK, Medline SKU # DYNJ57713C; 28. SPECIALS PK, Medline SKU # DYNJ57713D; 29. SPECIALS PK, Medline SKU # DYNJ57713F; 30. CARDIAC CATH LAB, Medline SKU # DYNJ57748A; 31. TAVR PACK, Medline SKU # DYNJ58771G; 32. CATH PACK GWV, Medline SKU # DYNJ59229F; 33. HEART CATH PACK, Medline SKU # DYNJ61294F; 34. CATH LAB PACK, Medline SKU # DYNJ61353C; 35. DEVICE IMPLANT PACK, Medline SKU # DYNJ64636A; 36. CUSTOM PROCEDURE KIT, Medline SKU # DYNJ65277A; 37. GENERAL VASCULAR PACK RFID-V2, Medline SKU # DYNJ66442F; 38. CARDIAC CATH PACK, Medline SKU # DYNJ69845C; 39. OMF MINOR PACK, Medline SKU # DYNJ83991C; 40. HEART CATH PACK, Medline SKU # DYNJ84133B; 41. SPECIAL PROCEDURES, Medline SKU # DYNJ85425B; 42. PRIMARY PACK, Medline SKU # DYNJT3093; 43. CARDIAC CATH PACK, Medline SKU # DYNJT3522A; 44. CARDIAC CATH PACK, Medline SKU # DYNJT3522B; 45. TOTAL KNEE PACK, Medline SKU # DYNJT4069; 46. CARDIAC CATH PACK, Medline SKU # DYNJT5685; 47. CATH PACK, Medline SKU # DYNJT6544; 48. PACK CARDIAC CATH, Medline SKU # DYNJT7591; 49. OBL CATH VEIN PACK, Medline SKU # DYNJT7863; 50. TUMESCENT SYRINGE KIT, Medline SKU # DYNJTUMSYR; 51. CSTM LH KIT, Medline SKU # VASC1054; 52. LHK, Medline SKU # VASC1130A; 53. MANIFOLD KIT, Medline SKU # VASC1137A; 54. LHK, Medline SKU # VASC1299A; 55. LHK, Medline SKU # VASC1399B; 56. LHK, Medline SKU # VASC1403; 57. IR 2 PORT, Medline SKU # VASC1404; 58. LHK, Medline SKU # VASC1405; 59. LHK, Medline SKU # VASC1455A; 60. TWO PORT MANIFOLD KIT, Medline SKU # VASC1496; 61. RHK, US, NAMIC MANIFOLD, Medline SKU # VASC1502; 62. LHK, Medline SKU # VASC1555; 63. 4 VALVE,PERC,MANIFOLD,KIT, Medline SKU # VASC1609; 64. LHK, Medline SKU # VASC1668; 65. LHK, Medline SKU # VASC1694; 66. LHK, Medline SKU # VASC1805; 67. TVR, Medline SKU # VASC1816.

Reason

Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Code Info

Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 26BBI625; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 26ABB718; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25JBJ140; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25HBF702; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25GBC747; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25DBR341; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25CBK686; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25ABR102; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 24IBK765; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 24GBS605; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 24EBO156; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 24BBG055; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 24ABH036; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 23LBI829; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 23KBN009; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 23FBO996; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 23FBA259; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 23EBD151; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 24IBE239; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 24HBR112; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 24GBY032; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 24BBO470; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 24ABB436; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 23LBA225; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 23GBK257; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 23FBG433; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 23FBC618; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 23EBX036; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 23DBD220; Medline SKU # 60131209, UDI/DI each 10193489046779, UDI/DI case 40193489046770, Lot Number: 24HBJ211; Medline SKU # 60160075, UDI/DI each 10193489048872, UDI/DI case 40193489048873, Lot Number: 24EBQ349; Medline SKU # 60160075, UDI/DI each 10193489048872, UDI/DI case 40193489048873, Lot Number: 24CBD613; Medline SKU # 60160075, UDI/DI each 10193489048872, UDI/DI case 40193489048873, Lot Number: 24ABL640; Medline SKU # 60160075, UDI/DI each 10193489048872, UDI/DI case 40193489048873, Lot Number: 23KBB857; Medline SKU # 60160075, UDI/DI each 10193489048872, UDI/DI case 40193489048873, Lot Number: 23IBC156; Medline SKU # 60160075, UDI/DI each 10193489048872, UDI/DI case 40193489048873, Lot Number: 23FBJ161; Medline SKU # 60160075, UDI/DI each 10193489048872, UDI/DI case 40193489048873, Lot Number: 23FBC907; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 26ABD974; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 25JBM213; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 25FBK284; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 25FBK967; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 24KBN942; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 24JBC074; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 24HBQ995; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 24GBP001; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 24CBF239; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 24ABJ769; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 23LBP098; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 23HBB863; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 23GBQ587; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 23FBG492; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 23CBR482; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 25CBJ809; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 25BBS609; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 25BBL037; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 24KBI829; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 24JBX172; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 24IBO369; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 24EBI926; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 24DBV041; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 24ABB294; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 23KBT232; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 23KBA553; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 23IBD723; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 23HBV742; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 23HBU186; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 23GBA723; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 23DBF014; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 26BLA621; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25LLA491; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25JLA720; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25JLA241; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25HLB071; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25GLA678; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25ELA960; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25ELA670; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25DLA414; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25BLA539; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25ALA504; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 24JLA106; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 24JLA071; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 24ILA631; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 24HLA635; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 24HLA514; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 24GLA105; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 24FLA889; Medline SKU # BIOB686, UDI/DI each 10193489684025, UDI/DI case 40193489684026, Lot Number: 25IMF801; Medline SKU # BIOB686, UDI/DI each 10193489684025, UDI/DI case 40193489684026, Lot Number: 24FMH284; Medline SKU # DYNJ0376145M, UDI/DI each 10889942941643, UDI/DI case 40889942941644, Lot Number: 24ILA248; Medline SKU # DYNJ0392706K, UDI/DI each 10195327588212, UDI/DI case 40195327588213, Lot Number: 25LBP577; Medline SKU # DYNJ0392706K, UDI/DI each 10195327588212, UDI/DI case 40195327588213, Lot Number: 25IBR953; Medline SKU # DYNJ0451426O, UDI/DI each 10195327698690, UDI/DI case 40195327698691, Lot Number: 25FBF110; Medline SKU # DYNJ0451426O, UDI/DI each 10195327698690, UDI/DI case 40195327698691, Lot Number: 25EBI930; Medline SKU # DYNJ0451426O, UDI/DI each 10195327698690, UDI/DI case 40195327698691, Lot Number: 25DBE958; Medline SKU # DYNJ0451426O, UDI/DI each 10195327698690, UDI/DI case 40195327698691, Lot Number: 24JBN523; Medline SKU # DYNJ0451426O, UDI/DI each 10195327698690, UDI/DI case 40195327698691, Lot Number: 24IBC823; Medline SKU # DYNJ0451426P, UDI/DI each 10198459386336, UDI/DI case 40198459386337, Lot Number: 26BBK812; Medline SKU # DYNJ0451426P, UDI/DI each 10198459386336, UDI/DI case 40198459386337, Lot Number: 26ABI992; Medline SKU # DYNJ0451426P, UDI/DI each 10198459386336, UDI/DI case 40198459386337, Lot Number: 26ABJ025; Medline SKU # DYNJ0451426P, UDI/DI each 10198459386336, UDI/DI case 40198459386337, Lot Number: 25KBD666; Medline SKU # DYNJ0451426P, UDI/DI each 10198459386336, UDI/DI case 40198459386337, Lot Number: 25JBR728; Medline SKU # DYNJ0451426P, UDI/DI each 10198459386336, UDI/DI case 40198459386337, Lot Number: 25JBR851; Medline SKU # DYNJ0451426P, UDI/DI each 10198459386336, UDI/DI case 40198459386337, Lot Number: 25GBS862; Medline SKU # DYNJ07953N, UDI/DI each 10198459123412, UDI/DI case 40198459123413, Lot Number: 25IBK486; Medline SKU # DYNJ07953N, UDI/DI each 10198459123412, UDI/DI case 40198459123413, Lot Number: 25IBF054; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 25KBI063; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 25IBK151; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 25EBD486; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 25DBL885; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 25CBL230; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24LBG685; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24KBO528; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24JBH940; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24HBQ365; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24HBJ343; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24GBE178; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24FBN018; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24EBR279; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24CBP560; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24ABP475; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 23LBS358; Medline SKU # DYNJ19343J, UDI/DI each 10888277827981, UDI/DI case 40888277827982, Lot Number: 24KBL564; Medline SKU # DYNJ20898T, UDI/DI each 10198459477003, UDI/DI case 40198459477004, Lot Number: 25KMH265; Medline SKU # DYNJ20898T, UDI/DI each 10198459477003, UDI/DI case 40198459477004, Lot Number: 25JML139; Medline SKU # DYNJ24250G, UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 26AMC371; Medline SKU # DYNJ24250G, UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 26AMC433; Medline SKU # DYNJ24250G, UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 25KBH544; Medline SKU # DYNJ24250G, UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 25JBG417; Medline SKU # DYNJ24250G, UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 25GBQ931; Medline SKU # DYNJ24250G, UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 25GBI702; Medline SKU # DYNJ38333D, UDI/DI each 10198459032981, UDI/DI case 40198459032982, Lot Number: 25AMG765; Medline SKU # DYNJ38333D, UDI/DI each 10198459032981, UDI/DI case 40198459032982, Lot Number: 24KMJ018; Medline SKU # DYNJ38333D, UDI/DI each 10198459032981, UDI/DI case 40198459032982, Lot Number: 24JME459; Medline SKU # DYNJ38333D, UDI/DI each 10198459032981, UDI/DI case 40198459032982, Lot Number: 24HMB740; Medline SKU # DYNJ38333D, UDI/DI each 10198459032981, UDI/DI case 40198459032982, Lot Number: 24GMH056; Medline SKU # DYNJ38333F, UDI/DI each 10198459258633, UDI/DI case 40198459258634, Lot Number: 25KME742; Medline SKU # DYNJ38333F, UDI/DI each 10198459258633, UDI/DI case 40198459258634, Lot Number: 25HME818; Medline SKU # DYNJ38333F, UDI/DI each 10198459258633, UDI/DI case 40198459258634, Lot Number: 25EMB706; Medline SKU # DYNJ38333F, UDI/DI each 10198459258633, UDI/DI case 40198459258634, Lot Number: 25CME621; Medline SKU # DYNJ38910D, UDI/DI each 10198459319815, UDI/DI case 40198459319816, Lot Number: 25GBJ993; Medline SKU # DYNJ38910D, UDI/DI each 10198459319815, UDI/DI case 40198459319816, Lot Number: 25EBD960; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 26ABR937; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25KBI764; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25JBR971; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25IBM264; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25IBI796; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25IBD184; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25FBV148; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25FBR538; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25FBB566; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25EBN741; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25EBF462; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25DBK395; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25CBU744; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25CBJ971; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25CBH568; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25ABK344; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24KBG845; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24IBM450; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24IBH804; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24HBL014; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24GBU989; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24EBQ673; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24EBO710; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24DBQ947; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24CBH211; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24BBE538; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24ABU829; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24ABA618; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 23KBO467; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 23IBS704; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 23IBM775; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 23HBX517; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 23GBO122; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 23FBW049; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 23EBG231; Medline SKU # DYNJ43403D, UDI/DI each 10195327179236, UDI/DI case 40195327179237, Lot Number: 23FLA639; Medline SKU # DYNJ48487F, UDI/DI each 10198459572937, UDI/DI case 40198459572938, Lot Number: 26AMA021; Medline SKU # DYNJ50779G, UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 25DMH709; Medline SKU # DYNJ50779G, UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 25CMI786; Medline SKU # DYNJ50779G, UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 25BMF563; Medline SKU # DYNJ50779G, UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 25BMC031; Medline SKU # DYNJ50779G, UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 24KMC197; Medline SKU # DYNJ50779G, UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 24IMI472; Medline SKU # DYNJ50779G, UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 24HMA312; Medline SKU # DYNJ50779G, UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 24GMA788; Medline SKU # DYNJ50779G, UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 24EMG321; Medline SKU # DYNJ50779I, UDI/DI each 10198459423758, UDI/DI case 40198459423759, Lot Number: 25LMC853; Medline SKU # DYNJ50779I, UDI/DI each 10198459423758, UDI/DI case 40198459423759, Lot Number: 25KMB250; Medline SKU # DYNJ50779I, UDI/DI each 10198459423758, UDI/DI case 40198459423759, Lot Number: 25IME495; Medline SKU # DYNJ50779I, UDI/DI each 10198459423758, UDI/DI case 40198459423759, Lot Number: 25HME177; Medline SKU # DYNJ50779I, UDI/DI each 10198459423758, UDI/DI case 40198459423759, Lot Number: 25GMF741; Medline SKU # DYNJ57049N, UDI/DI each 10198459519994, UDI/DI case 40198459519995, Lot Number: 25IMH467; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 24GBV109; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 24GBK784; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 24EBR407; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 24DBR592; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 24CBJ904; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 24ABS845; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 23KBR978; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 23JBX436; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 23IBJ923; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 23HBV971; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 23GBV284; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 23FBK372; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 23DBO863; Medline SKU # DYNJ57713D, UDI/DI each 10198459106583, UDI/DI case 40198459106584, Lot Number: 25DBT278; Medline SKU # DYNJ57713D, UDI/DI each 10198459106583, UDI/DI case 40198459106584, Lot Number: 25CBD829; Medline SKU # DYNJ57713D, UDI/DI each 10198459106583, UDI/DI case 40198459106584, Lot Number: 25BBU169; Medline SKU # DYNJ57713D, UDI/DI each 10198459106583, UDI/DI case 40198459106584, Lot Number: 25ABB803; Medline SKU # DYNJ57713D, UDI/DI each 10198459106583, UDI/DI case 40198459106584, Lot Number: 24KBS852; Medline SKU # DYNJ57713D, UDI/DI each 10198459106583, UDI/DI case 40198459106584, Lot Number: 24IBP601; Medline SKU # DYNJ57713D, UDI/DI each 10198459106583, UDI/DI case 40198459106584, Lot Number: 24IBA400; Medline SKU # DYNJ57713F, UDI/DI each 10198459380761, UDI/DI case 40198459380762, Lot Number: 26ABU685; Medline SKU # DYNJ57713F, UDI/DI each 10198459380761, UDI/DI case 40198459380762, Lot Number: 26ABS702; Medline SKU # DYNJ57713F, UDI/DI each 10198459380761, UDI/DI case 40198459380762, Lot Number: 25KBM524; Medline SKU # DYNJ57713F, UDI/DI each 10198459380761, UDI/DI case 40198459380762, Lot Number: 25JBD873; Medline SKU # DYNJ57713F, UDI/DI each 10198459380761, UDI/DI case 40198459380762, Lot Number: 25IBR638; Medline SKU # DYNJ57713F, UDI/DI each 10198459380761, UDI/DI case 40198459380762, Lot Number: 25IBB474; Medline SKU # DYNJ57713F, UDI/DI each 10198459380761, UDI/DI case 40198459380762, Lot Number: 25FBQ397; Medline SKU # DYNJ57748A, UDI/DI each 10888277746015, UDI/DI case 40888277746016, Lot Number: 24JME277; Medline SKU # DYNJ57748A, UDI/DI each 10888277746015, UDI/DI case 40888277746016, Lot Number: 24FMA227; Medline SKU # DYNJ57748A, UDI/DI each 10888277746015, UDI/DI case 40888277746016, Lot Number: 24EMG444; Medline SKU # DYNJ58771G, UDI/DI each 10195327663308, UDI/DI case 40195327663309, Lot Number: 25ALB024; Medline SKU # DYNJ58771G, UDI/DI each 10195327663308, UDI/DI case 40195327663309, Lot Number: 24KLA767; Medline SKU # DYNJ58771G, UDI/DI each 10195327663308, UDI/DI case 40195327663309, Lot Number: 24JLA065; Medline SKU # DYNJ58771G, UDI/DI each 10195327663308, UDI/DI case 40195327663309, Lot Number: 24ILA329; Medline SKU # DYNJ58771G, UDI/DI each 10195327663308, UDI/DI case 40195327663309, Lot Number: 24HLA502; Medline SKU # DYNJ58771G, UDI/DI each 10195327663308, UDI/DI case 40195327663309, Lot Number: 24FLA645; Medline SKU # DYNJ58771G, UDI/DI each 10195327663308, UDI/DI case 40195327663309, Lot Number: 24ELA005; Medline SKU # DYNJ58771G, UDI/DI each 10195327663308, UDI/DI case 40195327663309, Lot Number: 24DLA522; Medline SKU # DYNJ58771G, UDI/DI each 10195327663308, UDI/DI case 40195327663309, Lot Number: 24DLA423; Medline SKU # DYNJ59229F, UDI/DI each 10198459483936, UDI/DI case 40198459483937, Lot Number: 26BLA718; Medline SKU # DYNJ59229F, UDI/DI each 10198459483936, UDI/DI case 40198459483937, Lot Number: 25LLA775; Medline SKU # DYNJ59229F, UDI/DI each 10198459483936, UDI/DI case 40198459483937, Lot Number: 25ILA548; Medline SKU # DYNJ59229F, UDI/DI each 10198459483936, UDI/DI case 40198459483937, Lot Number: 25ILA434; Medline SKU # DYNJ59229F, UDI/DI each 10198459483936, UDI/DI case 40198459483937, Lot Number: 25ILA036; Medline SKU # DYNJ59229F, UDI/DI each 10198459483936, UDI/DI case 40198459483937, Lot Number: 25HLA821; Medline SKU # DYNJ61294F, UDI/DI each 10195327691158, UDI/DI case 40195327691159, Lot Number: 25HBP380; Medline SKU # DYNJ61353C, UDI/DI each 10198459503054, UDI/DI case 40198459503055, Lot Number: 25LMG882; Medline SKU # DYNJ61353C, UDI/DI each 10198459503054, UDI/DI case 40198459503055, Lot Number: 25KMB331; Medline SKU # DYNJ61353C, UDI/DI each 10198459503054, UDI/DI case 40198459503055, Lot Number: 25JMK949; Medline SKU # DYNJ64636A, UDI/DI each 10195327374617, UDI/DI case 40195327374618, Lot Number: 25FBE213; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 26ABC312; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 25LBG126; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 25IBS553; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 25IBN181; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 25GBT426; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 25FBP518; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 25BBR414; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 24KBO515; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 24JBT061; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 24GBY011; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 24EBS347; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 24EBS501; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 24CBS901; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 24BBD609; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 24BBD609A; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 23KBJ915; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 23JBH747; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 23HBF936; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 23GBA827; Medline SKU # DYNJ66442F, UDI/DI each 10198459334061, UDI/DI case 40198459334062, Lot Number: 25DBO052A; Medline SKU # DYNJ69845C, UDI/DI each 10198459657467, UDI/DI case 40198459657468, Lot Number: 26ABV680; Medline SKU # DYNJ83991C, UDI/DI each 10198459629945, UDI/DI case 40198459629946, Lot Number: 26AMF581; Medline SKU # DYNJ84133B, UDI/DI each 10198459501777, UDI/DI case 40198459501778, Lot Number: 25KBB504; Medline SKU # DYNJ85425B, UDI/DI each 10198459305566, UDI/DI case 40198459305567, Lot Number: 26BMH078; Medline SKU # DYNJ85425B, UDI/DI each 10198459305566, UDI/DI case 40198459305567, Lot Number: 25LMH057; Medline SKU # DYNJ85425B, UDI/DI each 10198459305566, UDI/DI case 40198459305567, Lot Number: 25JMD619; Medline SKU # DYNJ85425B, UDI/DI each 10198459305566, UDI/DI case 40198459305567, Lot Number: 25IMH487; Medline SKU # DYNJ85425B, UDI/DI each 10198459305566, UDI/DI case 40198459305567, Lot Number: 25IMD325; Medline SKU # DYNJ85425B, UDI/DI each 10198459305566, UDI/DI case 40198459305567, Lot Number: 25FMJ567; Medline SKU # DYNJ85425B, UDI/DI each 10198459305566, UDI/DI case 40198459305567, Lot Number: 25FMH927; Medline SKU # DYNJT3093, UDI/DI each 10198459305801, UDI/DI case 40198459305802, Lot Number: 25EBB099; Medline SKU # DYNJT3522A, UDI/DI each 10198459540479, UDI/DI case 40198459540470, Lot Number: 25JMI340; Medline SKU # DYNJT3522B, UDI/DI each 10198459602825, UDI/DI case 40198459602826, Lot Number: 25LMJ747; Medline SKU # DYNJT3522B, UDI/DI each 10198459602825, UDI/DI case 40198459602826, Lot Number: 25KME591; Medline SKU # DYNJT4069, UDI/DI each 10198459394751, UDI/DI case 40198459394752, Lot Number: 25JMI514; Medline SKU # DYNJT4069, UDI/DI each 10198459394751, UDI/DI case 40198459394752, Lot Number: 25GMB597; Medline SKU # DYNJT5685, UDI/DI each 10198459514098, UDI/DI case 40198459514099, Lot Number: 25JLA132; Medline SKU # DYNJT6544, UDI/DI each 10198459558689, UDI/DI case 40198459558680, Lot Number: 25KMF858; Medline SKU # DYNJT7591, UDI/DI each 10198459635823, UDI/DI case 40198459635824, Lot Number: 26AMH405; Medline SKU # DYNJT7863, UDI/DI each 10198459654831, UDI/DI case 40198459654832, Lot Number: 26ABT695; Medline SKU # DYNJTUMSYR, UDI/DI each 10884389882054, UDI/DI case 40884389882055, Lot Number: 26BBF460; Medline SKU # DYNJTUMSYR, UDI/DI each 10884389882054, UDI/DI case 40884389882055, Lot Number: 26ABT525; Medline SKU # DYNJTUMSYR, UDI/DI each 10884389882054, UDI/DI case 40884389882055, Lot Number: 25LBP400; Medline SKU # DYNJTUMSYR, UDI/DI each 10884389882054, UDI/DI case 40884389882055, Lot Number: 25LBK861; Medline SKU # DYNJTUMSYR, UDI/DI each 10884389882054, UDI/DI case 40884389882055, Lot Number: 25LBK862; Medline SKU # VASC1054, UDI/DI each 10889942379217, UDI/DI case 40889942379218, Lot Number: 25GBM922; Medline SKU # VASC1054, UDI/DI each 10889942379217, UDI/DI case 40889942379218, Lot Number: 25FBV097; Medline SKU # VASC1130A, UDI/DI each 10198459040092, UDI/DI case 40198459040093, Lot Number: 25HBE924; Medline SKU # VASC1137A, UDI/DI each 10195327636692, UDI/DI case 40195327636693, Lot Number: 25HBU244; Medline SKU # VASC1299A, UDI/DI each 10195327076450, UDI/DI case 40195327076451, Lot Number: 25FBQ566; Medline SKU # VASC1299A, UDI/DI each 10195327076450, UDI/DI case 40195327076451, Lot Number: 25FBH292; Medline SKU # VASC1299A, UDI/DI each 10195327076450, UDI/DI case 40195327076451, Lot Number: 24HBE146; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 26ABT499; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 25LBD803; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 25IBN891; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 25GBM607; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 25FBI750; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 25EBQ507; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 25DBT209; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 25BBS353; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 24KBQ595; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 24HBX234; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 24EBT407; Medline SKU # VASC1403, UDI/DI each 10195327366285, UDI/DI case 40195327366286, Lot Number: 24BBU522; Medline SKU # VASC1403, UDI/DI each 10195327366285, UDI/DI case 40195327366286, Lot Number: 23KBJ722; Medline SKU # VASC1404, UDI/DI each 10195327366568, UDI/DI case 40195327366569, Lot Number: 25DBE744; Medline SKU # VASC1404, UDI/DI each 10195327366568, UDI/DI case 40195327366569, Lot Number: 25FBH475; Medline SKU # VASC1405, UDI/DI each 10195327367015, UDI/DI case 40195327367016, Lot Number: 25DBJ392; Medline SKU # VASC1455A, UDI/DI each 10198459025396, UDI/DI

More Code Info

case 40198459025397, Lot Number: 25IBD178; Medline SKU # VASC1496, UDI/DI each 10195327615666, UDI/DI case 40195327615667, Lot Number: 25FBH317; Medline SKU # VASC1502, UDI/DI each 10195327667719, UDI/DI case 40195327667710, Lot Number: 24GBJ600; Medline SKU # VASC1502, UDI/DI each 10195327667719, UDI/DI case 40195327667710, Lot Number: 24FBL559; Medline SKU # VASC1555, UDI/DI each 10198459058288, UDI/DI case 40198459058289, Lot Number: 25HBP433; Medline SKU # VASC1555, UDI/DI each 10198459058288, UDI/DI case 40198459058289, Lot Number: 25HBG954; Medline SKU # VASC1555, UDI/DI each 10198459058288, UDI/DI case 40198459058289, Lot Number: 25GBH483; Medline SKU # VASC1555, UDI/DI each 10198459058288, UDI/DI case 40198459058289, Lot Number: 25FBT540; Medline SKU # VASC1609, UDI/DI each 10198459144653, UDI/DI case 40198459144654, Lot Number: 25IBQ779; Medline SKU # VASC1668, UDI/DI each 10198459228155, UDI/DI case 40198459228156, Lot Number: 25JBK828; Medline SKU # VASC1668, UDI/DI each 10198459228155, UDI/DI case 40198459228156, Lot Number: 25IBG375; Medline SKU # VASC1668, UDI/DI each 10198459228155, UDI/DI case 40198459228156, Lot Number: 25EBG231; Medline SKU # VASC1668, UDI/DI each 10198459228155, UDI/DI case 40198459228156, Lot Number: 25EBE966; Medline SKU # VASC1668, UDI/DI each 10198459228155, UDI/DI case 40198459228156, Lot Number: 25BBE722; Medline SKU # VASC1694, UDI/DI each 10198459252501, UDI/DI case 40198459252502, Lot Number: 25HBN801; Medline SKU # VASC1805, UDI/DI each 10198459399022, UDI/DI case 40198459399023, Lot Number: 25HMI038; Medline SKU # VASC1816, UDI/DI each 10198459403224, UDI/DI case 40198459403225, Lot Number: 26BMF266; Medline SKU # VASC1816, UDI/DI each 10198459403224, UDI/DI case 40198459403225, Lot Number: 25JBQ974; Medline SKU # VASC1816, UDI/DI each 10198459403224, UDI/DI case 40198459403225, Lot Number: 25HBN802.

Distribution

Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).

Quantity

79843 units