Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: MINOR PLASTIC-LF, Medline Kit SKU # DYNJ86596
Enforcement
- Recall Number
- Z-1723-2026
- Event ID
- 98598
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medline Industries, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- April 15, 2026
- Initiation Date
- February 27, 2026
- Classification Date
- April 7, 2026
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: MINOR PLASTIC-LF, Medline Kit SKU # DYNJ86596
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Medline Kit SKU # DYNJ86596, UDI/DI each 10195327656119, UDI/DI case 40195327656110, Lot Numbers: 24FMI029, 24DMK138.
Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).
60 kits