REF 154232 Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Size E7 ARCOM UHMWPE / CO-CR-MO Alloy Only for use in Lateral compartment Sterile R The Oxford Fixed Bearing Knee is a Partial knee replacement system consisting of a femoral component and a fixed bearing tibial component for use in the lateral compartment of the knee. The Oxford Fixed Lateral bearings are indicated for use in the lateral compartment and intended to be implemented with bone cement.
Enforcement
- Recall Number
- Z-1719-2013
- Event ID
- 65459
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Biomet, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 17, 2013
- Initiation Date
- June 7, 2013
- Classification Date
- July 11, 2013
- Termination Date
- July 17, 2014
- Address
- 56 E Bell Dr, N/A, Warsaw, IN, 46582-6989, United States
Description
REF 154232 Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Size E7 ARCOM UHMWPE / CO-CR-MO Alloy Only for use in Lateral compartment Sterile R The Oxford Fixed Bearing Knee is a Partial knee replacement system consisting of a femoral component and a fixed bearing tibial component for use in the lateral compartment of the knee. The Oxford Fixed Lateral bearings are indicated for use in the lateral compartment and intended to be implemented with bone cement.
Biomet investigation found the design of the tibial bearing creates the potential for anterior impingement between the bone and the poly bearing component in terminal extension.
M921180, M539440, M444920, M722450, M136990, M657080
Worldwide distribution: USA including NC, IN, PA, and OH; and countries of: Canada and Netherlands.
31