FDA Enforcement Class II Terminated

Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers.

Recall: Z-1717-2020 · Reported April 22, 2020

Enforcement

Recall Number
Z-1717-2020
Event ID
84759
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Sysmex America, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 22, 2020
Initiation Date
January 21, 2020
Classification Date
April 15, 2020
Termination Date
January 30, 2023
Address
1 Sysmex Way, Mundelein, IL, 60060-9528, United States

Description

Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers.

Reason

When excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a combustion hazard.

Code Info

Serial Numbers: 11201, 11202, 12310, 12311, 12312, 12313, 12941, 12942, 12997, 13875, 14170, 14171, 14172, 14173, 14174, 14175, 14278, 14279, 14280, 14962, 14963, 14964, 14965, 14966, 14967, 14968, 14969, 14970, 14972, 14973, 14974, 14975, 14976, 14977, 14978, 14979, 15017, 15018, 15619, 15620

Distribution

USA: FL, VT, IL.

Quantity

40 units