FDA Enforcement
Class II
Terminated
The NAV3i Platform Power Box supplies the electrical components with power (e.g. Stryker Nav3 Graphite Camera, SPC-3 Computer, IO-Tablet (touch screen) and the monitor).
Recall: Z-1716-2014
·
Reported June 11, 2014
Enforcement
- Recall Number
- Z-1716-2014
- Event ID
- 68261
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Stryker Instruments Div. of Stryker Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Report Date
- June 11, 2014
- Initiation Date
- May 13, 2014
- Classification Date
- June 4, 2014
- Termination Date
- September 15, 2014
- Address
- 4100 E Milham Ave, Portage, MI, 49002-9704, United States
Description
The NAV3i Platform Power Box supplies the electrical components with power (e.g. Stryker Nav3 Graphite Camera, SPC-3 Computer, IO-Tablet (touch screen) and the monitor).
Reason
The Stryker NAV3i Platform Power Boxes produced between December 9, 2013 and March 7, 2014 may have been assembled without electrical insulation of the soldering joints
Code Info
Stryker Product Number: 7700-853-000 Stryker Serial Numbers: 10004
Distribution
Distributed in the states of AR, CA, IN, LA, MA, MD, MI, MT, NV, and TX.
Quantity
20