Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. CSTM MANIFOLD KIT - TITUS, Medline SKU VASC1061; 2. CSTM MANIFOLD KIT - FRYE, Medline SKU VASC1074; 3. CRYO EP MANIFOLD KIT, Medline SKU VASC1198; 4. MANIFOLD KIT - 500PSI OFF, Medline SKU VASCSHPOFF1; 5. MANIFOLD KIT LOW PRESSURE ON, Medline SKU VASCSLPON1.
Enforcement
- Recall Number
- Z-1715-2026
- Event ID
- 98598
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medline Industries, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- April 15, 2026
- Initiation Date
- February 27, 2026
- Classification Date
- April 7, 2026
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. CSTM MANIFOLD KIT - TITUS, Medline SKU VASC1061; 2. CSTM MANIFOLD KIT - FRYE, Medline SKU VASC1074; 3. CRYO EP MANIFOLD KIT, Medline SKU VASC1198; 4. MANIFOLD KIT - 500PSI OFF, Medline SKU VASCSHPOFF1; 5. MANIFOLD KIT LOW PRESSURE ON, Medline SKU VASCSLPON1.
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Medline SKU VASC1061, UDI/DI each 10889942464722, UDI/DI case 40889942464723, Lot Number 25IBN161; Medline SKU VASC1061, UDI/DI each 10889942464722, UDI/DI case 40889942464723, Lot Number 25GBF769; Medline SKU VASC1074, UDI/DI each 10889942611751, UDI/DI case 40889942611752, Lot Number 25HBN798; Medline SKU VASC1074, UDI/DI each 10889942611751, UDI/DI case 40889942611752, Lot Number 25GBL527; Medline SKU VASC1198, UDI/DI each 10193489664263, UDI/DI case 40193489664264, Lot Number 25GDA142; Medline SKU VASCSHPOFF1, UDI/DI each 10889942405503, UDI/DI case 40889942405504, Lot Number 25IBP989; Medline SKU VASCSHPOFF1, UDI/DI each 10889942405503, UDI/DI case 40889942405504, Lot Number 25GBL526; Medline SKU VASCSLPON1, UDI/DI each 10889942405510, UDI/DI case 40889942405511, Lot Number 25GBL070.
Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).
1698 kits