FDA Enforcement Class II Terminated

Access 2 assay Immunoassay System, Remanufactured, Part No. 386220 Product Usage: The Access 2 and UniCel DxI Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.

Recall: Z-1710-2017 · Reported April 12, 2017

Enforcement

Recall Number
Z-1710-2017
Event ID
76688
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Beckman Coulter Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 12, 2017
Initiation Date
March 7, 2017
Classification Date
April 4, 2017
Termination Date
March 1, 2018
Address
250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States

Description

Access 2 assay Immunoassay System, Remanufactured, Part No. 386220 Product Usage: The Access 2 and UniCel DxI Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.

Reason

Beckman Coulter is recalling the Access 2 Family of Immunoassay Systems because a circuit board that controls motor functions can affect device positioning on the pipettor and incubator belt while the instrument is running. The position errors cause the instrument to stop running and cancel tests that are in process.

Code Info

Serial No. 501012,501110,501190,501224,501294,501371,501394,501403,501440, 501443,501490,501501,501673,501681,501693,501861,501920,501926, 501936,501970,502165,502196,502319,502331,502332,502368,502458, 502475,502506,502543,502549,502556,502574,502640,502647,502651, 502791,502843,502848,503153,503279,503313,503354,503384,503392, 503424,503543,503627,503678,504046,504071,504072,504092,504112, 504195,504322,504365,504370,504756,504767,504902,505042,505195, 505251,505254,505644,505795,505840,505947,506002,506026,506108, 506119,506153,506187,506371,506406,506560,506586,506664,506708, 506766,506786,506837,506846,507039,507120,507161,507162,507214, 507229,507248,507282,507353,507397,507480,507511,507604,507619, 507842,507929,508019,508096,508179,508227,508313,508334,508412, 508478,508509,508516,508767,509051,509087,509113,509164,509253, 509297,509632,509837,510049,511042,511047,511180,511383,700024, 800321,800341,800443,800608,800611,800650,800849

Distribution

Affected geographies: Algeria, Austria, Australia, Bangladesh, Bulgaria, Brazil, Canada, China, Costa Rica, Cyprus, Colombia, Czech Republic, El Salvador, France, Germany, Ghana, Greece, Guatemala, India, Indonesia, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Lithuania, Malawi, Mexico, Morocco, Myanmar, Nepal, Oman, Panama, Philippines, Pakistan, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Serbia, South Africa, South Korea, Spain, Switzerland, Tunisia, Turkey, United Kingdom, United States, Vietnam, Zimbabwe

Quantity

851 units total for all products (291 units in US)