FDA Enforcement Class II Ongoing

Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): RPC-35-180 ORDER NUMBER (GPN):G07518. For vascular catheter positioning and exchange in diagnostic and interventional procedures.

Recall: Z-1709-2022 · Reported September 21, 2022

Enforcement

Recall Number
Z-1709-2022
Event ID
90710
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Cook Incorporated
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 21, 2022
Initiation Date
July 19, 2022
Classification Date
September 9, 2022
Address
750 N Daniels Way, Bloomington, IN, 47404-9120, United States

Description

Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): RPC-35-180 ORDER NUMBER (GPN):G07518. For vascular catheter positioning and exchange in diagnostic and interventional procedures.

Reason

Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.

Code Info

UDI-DI: (01)00827002075184 Lot Numbers: 14063361, 14087089, 14087092, 14117033, 14120926, 14120928, 14151216, 14151218, 14153159, 14158614, 14159695, 14159696

Distribution

Nationwide Foreign: AR AT AU BE BG CA CA CH CL CN CZ DE DK ES FR GB GR GU HU IL IT JO JP KR KW MX MY NC NL PE PL PR PT PY RU SE SI TR UY ZA

Quantity

1997 units