FDA Enforcement Class II Ongoing

IPS e.max ZirCAD CER/in. Prime C2 C17/5- A zirconium oxide block for the fabrication of fixed, full-contour restorations in anterior and posterior teeth. Article Number: 758443

Recall: Z-1708-2024 · Reported May 8, 2024

Enforcement

Recall Number
Z-1708-2024
Event ID
94401
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
IVOCLAR VIVADENT AG
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
May 8, 2024
Initiation Date
April 5, 2024
Classification Date
April 30, 2024
Address
BEDERSRASSE, LIECHTSTEIN, N/A, Germany

Description

IPS e.max ZirCAD CER/in. Prime C2 C17/5- A zirconium oxide block for the fabrication of fixed, full-contour restorations in anterior and posterior teeth. Article Number: 758443

Reason

Defective blocks may show cracks after sintering If cracks are unnoticed, restorations may fracture in the patient's mouth and require restoration

Code Info

UDI-DI: (01)07615208438611 (11)231130 (10)Z065M1 (241)758443 Batch No.: Z065M1

Distribution

US Nationwide distribution.

Quantity

2472 units