FDA Enforcement Class II Terminated

Stryker Neurovascular, AXS Infinity LS Plus Long Sheath, Rx Only, Sterile EO, USA Customer Service 855-91 NEURO (916-3876): a) REF: INC-11196-70, 70cm, b) REF: INC-11196-80, 80cm, c) REF: INC-11196-90, 90cm Product Usage: Indications for Use: The AXS Infinity LS Plus Long Sheath is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. Device Description: The AXS Infinity LS Plus Long Sheath is a single lumen, flexible, variable stiffness long sheath with an 0.091 inch inner diameter. It has a radiopaque marker band on the distal end and a Luer hub at the proximal end. The AXS Infinity LS Plus Long Sheath shaft has a 10 cm lubricious coating at the distal end to reduce friction during use. The inner lumen of the AXS Infinity LS Plus Long Sheath is compatible with 6F or smaller catheters. Each package includes one AXS Infinity LS Plus Long Sheath (INC- 11196-70, INC-11196-80, or INC-11196-90), one Dilator, and one hemostasis valve. Dimensions of the AXS Infinity LS Plus Long Sheath are included on the individual device label.

Recall: Z-1708-2019 · Reported June 12, 2019

Enforcement

Recall Number
Z-1708-2019
Event ID
82755
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Neurovascular
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 12, 2019
Initiation Date
April 18, 2019
Classification Date
June 2, 2019
Termination Date
April 27, 2021
Address
47900 Bayside Pkwy, Fremont, CA, 94538-6515, United States

Description

Stryker Neurovascular, AXS Infinity LS Plus Long Sheath, Rx Only, Sterile EO, USA Customer Service 855-91 NEURO (916-3876): a) REF: INC-11196-70, 70cm, b) REF: INC-11196-80, 80cm, c) REF: INC-11196-90, 90cm Product Usage: Indications for Use: The AXS Infinity LS Plus Long Sheath is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. Device Description: The AXS Infinity LS Plus Long Sheath is a single lumen, flexible, variable stiffness long sheath with an 0.091 inch inner diameter. It has a radiopaque marker band on the distal end and a Luer hub at the proximal end. The AXS Infinity LS Plus Long Sheath shaft has a 10 cm lubricious coating at the distal end to reduce friction during use. The inner lumen of the AXS Infinity LS Plus Long Sheath is compatible with 6F or smaller catheters. Each package includes one AXS Infinity LS Plus Long Sheath (INC- 11196-70, INC-11196-80, or INC-11196-90), one Dilator, and one hemostasis valve. Dimensions of the AXS Infinity LS Plus Long Sheath are included on the individual device label.

Reason

A design related issue can contribute to the device exhibiting brittle behavior. This behavior may result in fractures of the catheter shaft.

Code Info

All lots

Distribution

U.S Nationwide Distribution - PA, VA, TX, CA, FL, AR, TN, MI, NY, IL, NC, KY, WV, OH, MO, AZ, GA, WA, VT, DC, SC, NJ, MN, OK, MD, AL, LA, MA, DE, IN, OR, RI, SD, MS, WI, ID, UT, CO, IA, KS, CT, NE, NM

Quantity

2435