FDA Enforcement
Class II
Ongoing
Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): RFSPC-35-180 ORDER NUMBER (GPN) G09608. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
Recall: Z-1707-2022
·
Reported September 21, 2022
Enforcement
- Recall Number
- Z-1707-2022
- Event ID
- 90710
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Cook Incorporated
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 21, 2022
- Initiation Date
- July 19, 2022
- Classification Date
- September 9, 2022
- Address
- 750 N Daniels Way, Bloomington, IN, 47404-9120, United States
Description
Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): RFSPC-35-180 ORDER NUMBER (GPN) G09608. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
Reason
Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.
Code Info
UDI-DI: (01)00827002096080- Lot Numbers: 14071906, 14091204, 14134663, 14134667, 14134669, 14134670, 14134673, 14134676, 14143872, 14143873, 14160997, 14184102, NS14143735, NS14152903
Distribution
Nationwide Foreign: AR AT AU BE BG CA CA CH CL CN CZ DE DK ES FR GB GR GU HU IL IT JO JP KR KW MX MY NC NL PE PL PR PT PY RU SE SI TR UY ZA
Quantity
2407 units