Medline convenience kits containing Olympus biopsy valves: 1. BRONCH PACK, DYKE2096 2. NON MB BRONCHOPSCOPY CCH LH PK, DYNJT6182 3. PULMONARY BRONCH PACK, DYNDA2861A
Enforcement
- Recall Number
- Z-1706-2026
- Event ID
- 98530
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medline Industries, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Report Date
- April 8, 2026
- Initiation Date
- February 16, 2026
- Classification Date
- April 1, 2026
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Medline convenience kits containing Olympus biopsy valves: 1. BRONCH PACK, DYKE2096 2. NON MB BRONCHOPSCOPY CCH LH PK, DYNJT6182 3. PULMONARY BRONCH PACK, DYNDA2861A
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
1. UDI-DI 10198459629334 (EA) 40198459629335 (CS) LOTS 26AMG895 2. UDI-DI 10198459545283 (EA) 40198459545284 (CS) LOTS 25LBJ366 26ABR365 3. UDI-DI 10195327471859 (EA) 40195327471850 (CS) LOTS 24EBA317 24EBV431 24IBD694 24JBQ783
US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.
14,379 kits total