FDA Enforcement
Class II
Terminated
BEQ-T 9410 Lurie Chldns Small 1/4, Custom Tubing Kit, Catalog No. 709000430
Recall: Z-1706-2018
·
Reported May 16, 2018
Enforcement
- Recall Number
- Z-1706-2018
- Event ID
- 79873
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Datascope Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 16, 2018
- Initiation Date
- October 19, 2017
- Classification Date
- May 8, 2018
- Termination Date
- April 24, 2020
- Address
- 15 Law Dr, N/A, Fairfield, NJ, 07004-3206, United States
Description
BEQ-T 9410 Lurie Chldns Small 1/4, Custom Tubing Kit, Catalog No. 709000430
Reason
The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.
Code Info
Lot 3000060035
Distribution
Distributed to accounts in AL, AR, CA, DC, FL, IL, LA, MO, NC, PA, SC, TX, and WA.
Quantity
6