FDA Enforcement Class II Terminated

Spacelabs Medical Ultraview SL Command Module, Model 91496. Arrhythmia Detector and Alarm. The Multi-parameter Module is used with the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from a patient in a clinical environment.

Recall: Z-1705-2014 · Reported June 11, 2014

Enforcement

Recall Number
Z-1705-2014
Event ID
68357
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Spacelabs Healthcare Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Visit
Report Date
June 11, 2014
Initiation Date
May 8, 2014
Classification Date
June 3, 2014
Termination Date
June 3, 2014
Address
35301 SE Center St, Snoqualmie, WA, 98065-9216, United States

Description

Spacelabs Medical Ultraview SL Command Module, Model 91496. Arrhythmia Detector and Alarm. The Multi-parameter Module is used with the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from a patient in a clinical environment.

Reason

A software anomaly causes temperature alarm limits to change to limits which are not physiologically possible and result in nuisance false positive alarms.

Code Info

All devices of Model: 91496.

Distribution

US Distribution to GA.

Quantity

64 devices