FDA Enforcement
Class II
Terminated
Spacelabs Medical Ultraview SL Command Module, Model 91496. Arrhythmia Detector and Alarm. The Multi-parameter Module is used with the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from a patient in a clinical environment.
Recall: Z-1705-2014
·
Reported June 11, 2014
Enforcement
- Recall Number
- Z-1705-2014
- Event ID
- 68357
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Spacelabs Healthcare Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Visit
- Report Date
- June 11, 2014
- Initiation Date
- May 8, 2014
- Classification Date
- June 3, 2014
- Termination Date
- June 3, 2014
- Address
- 35301 SE Center St, Snoqualmie, WA, 98065-9216, United States
Description
Spacelabs Medical Ultraview SL Command Module, Model 91496. Arrhythmia Detector and Alarm. The Multi-parameter Module is used with the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from a patient in a clinical environment.
Reason
A software anomaly causes temperature alarm limits to change to limits which are not physiologically possible and result in nuisance false positive alarms.
Code Info
All devices of Model: 91496.
Distribution
US Distribution to GA.
Quantity
64 devices