FDA Enforcement Class II Ongoing

Abbott Cholestech LDX Battery Kit, Model: 3LP656194-PCM-LD, PN: 55170

Recall: Z-1702-2024 · Reported May 8, 2024

Enforcement

Recall Number
Z-1702-2024
Event ID
94190
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Alere San Diego, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 8, 2024
Initiation Date
March 1, 2024
Classification Date
April 30, 2024
Address
9975 Summers Ridge Rd, San Diego, CA, 92121-2997, United States

Description

Abbott Cholestech LDX Battery Kit, Model: 3LP656194-PCM-LD, PN: 55170

Reason

There is a potential that the Cholestech LDX battery (an external battery pack available as an optional accessory to the Cholestech LDX Analyzer) may swell which could cause minor burns, cuts or abrasions, or low current electrical shock to the user.

Code Info

All Lots/ UDI:

Distribution

Worldwide distribution - US Nationwide and the countries of United Kingdom, Netherlands and Kenya.

Quantity

559 units