FDA Enforcement Class II Terminated

Hologic M-IV Mammography System; Model M4--00001. Intended to produce radiographic images of the breast. For screening and diagnostic imaging of the breast.

Recall: Z-1701-2014 · Reported June 11, 2014

Enforcement

Recall Number
Z-1701-2014
Event ID
68298
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Hologic, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 11, 2014
Initiation Date
May 13, 2014
Classification Date
June 3, 2014
Termination Date
August 18, 2014
Address
36 Apple Ridge Rd # 37, Danbury, CT, 06810-7301, United States

Description

Hologic M-IV Mammography System; Model M4--00001. Intended to produce radiographic images of the breast. For screening and diagnostic imaging of the breast.

Reason

Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury

Code Info

Serial Numbers: 1800314 7263, 180031472, 18002023051PR

Distribution

Distributed in Puerto Rico, and the states of CA, NY, and TX, and the country of Republic of Korea.

Quantity

3 units