Terminated
Hologic M-IV Mammography System; Model M4--00001. Intended to produce radiographic images of the breast. For screening and diagnostic imaging of the breast.
Recall: Z-1701-2014
·
Reported June 11, 2014
Enforcement
- Recall Number
- Z-1701-2014
- Event ID
- 68298
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Hologic, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 11, 2014
- Initiation Date
- May 13, 2014
- Classification Date
- June 3, 2014
- Termination Date
- August 18, 2014
- Address
- 36 Apple Ridge Rd # 37, Danbury, CT, 06810-7301, United States
Description
Hologic M-IV Mammography System; Model M4--00001. Intended to produce radiographic images of the breast. For screening and diagnostic imaging of the breast.
Reason
Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury
Code Info
Serial Numbers: 1800314 7263, 180031472, 18002023051PR
Distribution
Distributed in Puerto Rico, and the states of CA, NY, and TX, and the country of Republic of Korea.
Quantity
3 units