FDA Enforcement Class II Terminated

Merlin PCS 3650 Software Upgrade Kit, REF 3330 St. Jude Medical The MerlinTM Patient Care System (MerlinTM PCS) Model 3650 is a portable, dedicated programming system designed to interrogate, program, display data, and test St. Jude MedicalTM implantable devices and leads.

Recall: Z-1700-2020 · Reported April 22, 2020

Enforcement

Recall Number
Z-1700-2020
Event ID
85275
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Abbott
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 22, 2020
Initiation Date
February 28, 2020
Classification Date
April 10, 2020
Termination Date
April 18, 2024
Address
15900 Valley View Ct, Sylmar, CA, 91342-3577, United States

Description

Merlin PCS 3650 Software Upgrade Kit, REF 3330 St. Jude Medical The MerlinTM Patient Care System (MerlinTM PCS) Model 3650 is a portable, dedicated programming system designed to interrogate, program, display data, and test St. Jude MedicalTM implantable devices and leads.

Reason

Internal testing of software identified software anomaly scenarios where if non-applicable shock energy value is selected, the device will continue to use the previously programmed energy level and the user will not be alerted that the selected value was not retained.

Code Info

Software Version: 24.6.1 Model: 3650, GTIN: 05414734500661 Model: 3330, GTIN: 05414734509725 For affected serial number of programmers see Attachment B.1 Programmers For models of ICDs interrogated by affected programmers see Attachment B.2 Devices.

Distribution

U.S. AK, Al, AR, AZ, CA, CO, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. OUS: None - the firm stated that there are no OUS consignees associated to ICDs that were interrogated from January 22, 2020 to February 6, 2020.

Quantity

Software Version 24.6.1 installed into 4733 programmers