FDA Enforcement Class II Terminated

Hologic Selenia Full Field Digital Mammography System: SEL-00002, SEL-00013, SEL-00030, SEL-00039. The Selenia Full Field Digital Mammography system generates digital mammography images that can be used for screening and diagnosis of breast cancer.

Recall: Z-1700-2014 · Reported June 11, 2014

Enforcement

Recall Number
Z-1700-2014
Event ID
68298
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Hologic, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 11, 2014
Initiation Date
May 13, 2014
Classification Date
June 3, 2014
Termination Date
August 18, 2014
Address
36 Apple Ridge Rd # 37, Danbury, CT, 06810-7301, United States

Description

Hologic Selenia Full Field Digital Mammography System: SEL-00002, SEL-00013, SEL-00030, SEL-00039. The Selenia Full Field Digital Mammography system generates digital mammography images that can be used for screening and diagnosis of breast cancer.

Reason

Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury

Code Info

Serial Numbers: 2840314W8065, 2840314W8062W, 2840314W8066W, 2840314W8067 2840314W8073, 2840314W8064, 2840314W8056W, 28401082720 28407083579, 28401137611W, 28403083045W

Distribution

Distributed in Puerto Rico, and the states of CA, NY, and TX, and the country of Republic of Korea.

Quantity

11 units