STA - Deficient VIII, Product Code: 00725. STA - Deficient VIII is an immuno-depleted human plasma intended for use in tests for the determination of factor VIII activity in plasma on STA-R and STA Compact analyzer platforms as follows: (A) STA-R Max, Product Codes: 58989, 58990 (B) STA-R Evolution, Product Codes: 58986, 58987, 58988 (C) STA Compact Max, Product Codes: 59026, 59027 (D) STA Compact, Product Codes: 58978, 58979
Enforcement
- Recall Number
- Z-1699-2020
- Event ID
- 85167
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Diagnostica Stago, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- April 22, 2020
- Initiation Date
- March 4, 2020
- Classification Date
- April 10, 2020
- Termination Date
- November 10, 2021
- Address
- 5 Century Dr, Parsippany, NJ, 07054-4607, United States
Description
STA - Deficient VIII, Product Code: 00725. STA - Deficient VIII is an immuno-depleted human plasma intended for use in tests for the determination of factor VIII activity in plasma on STA-R and STA Compact analyzer platforms as follows: (A) STA-R Max, Product Codes: 58989, 58990 (B) STA-R Evolution, Product Codes: 58986, 58987, 58988 (C) STA Compact Max, Product Codes: 59026, 59027 (D) STA Compact, Product Codes: 58978, 58979
A risk of sample-to-sample cross-contamination of Emicizumab for factor VIII assays on Stago instruments, clotting times will be shortened that may cause factor VIII levels reported to be falsely increased
All lots This cleaning sequence has been implemented in the following available software versions. Customers not possessing these software levels will be upgraded by Stago Field Support personnel: STA-R¿ Max (all serial numbers): Software version 4.05.04 or higher, and Methodology Update (MU) to enhance database version MS215. STA-R¿ Evolution (serial numbers higher than 8012756): Software version 3.04.09 or higher STA¿ Compact Max (all serial numbers): Software version 110.05 or higher.
Nationwide
3,225 units