FDA Enforcement
Class II
Terminated
Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (NCL-L-GSTpi-438), a liquid tissue culture supernatant. For in vitro diagnostic use.
Recall: Z-1698-2014
·
Reported June 11, 2014
Enforcement
- Recall Number
- Z-1698-2014
- Event ID
- 68065
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Leica Microsystems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 11, 2014
- Initiation Date
- February 26, 2014
- Classification Date
- June 2, 2014
- Termination Date
- August 14, 2017
- Address
- 1700 Leider Ln, Buffalo Grove, IL, 60089-6622, United States
Description
Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (NCL-L-GSTpi-438), a liquid tissue culture supernatant. For in vitro diagnostic use.
Reason
Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (Product Code: NCL-L-GSTpi-438) does not function as intended up to the expiry date on the product labeling. This was detected through an ongoing stability program. There is a linkage between the age of the product and staining intensity.
Code Info
Product Code: NCL-L-GSTpi-438; Lot Numbers and Expiration Dates: 6018777, Expires 2014-05; 6023260, Expires 2014-12; 6026297, Expires 2015-05; 6026980, Expires 2015-06
Distribution
Nationwide Distribution including AZ, CA, FL, and NY.
Quantity
33 units