FDA Enforcement Class I Ongoing

Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit Model/Catalog Number: AH165 Software Version: N/A Product Description: AirLife Infant Heated Wire Circuit dual-limb, dual-heat, high-flow circuit (>4 L/min) Component: No. a conduit for respiratory gas between the patient and a ventilator.

Recall: Z-1697-2025 · Reported May 21, 2025

Enforcement

Recall Number
Z-1697-2025
Event ID
96664
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Vyaire Medical
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 21, 2025
Initiation Date
April 10, 2025
Classification Date
May 9, 2025
Address
26125 N Riverwoods Blvd, Mettawa, IL, 60045-3420, United States

Description

Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit Model/Catalog Number: AH165 Software Version: N/A Product Description: AirLife Infant Heated Wire Circuit dual-limb, dual-heat, high-flow circuit (>4 L/min) Component: No. a conduit for respiratory gas between the patient and a ventilator.

Reason

Adapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia, and organ failure.

Code Info

Lot Code: Lot/Serial Number(s): 0004240347 0004240348 0004252021 0004253194 0004253470 0004255176 0004260100 0004262183 0004262987 0004263371 0004292077 0004300092 0004301668 UDI-DI AirLife Label Each: 10889483595862 Case: 30889483595866 Vyaire Label Each: 10190752145139 Case: 50190752145137

Distribution

Worldwide distribution - US Nationwide and the countries of Germany, Poland, Belgium, Spain, Romania, UK, Italy, Netherlands, India, Slovenia, Qatar, and Saudi Arabia.

Quantity

1895