FDA Enforcement Class II Terminated

CyberKnife VSI, Part Numbers: a) 032500-010: CyberKnife VSI 9.6.x Robotic Radiosurgery System for Japan b) 033000-010: CyberKnife VSI Robotic Radiosurgery System

Recall: Z-1697-2019 · Reported June 12, 2019

Enforcement

Recall Number
Z-1697-2019
Event ID
82664
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Accuray Incorporated
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 12, 2019
Initiation Date
April 10, 2018
Classification Date
May 31, 2019
Termination Date
April 1, 2021
Address
1240 Deming Way, Madison, WI, 53717-1954, United States

Description

CyberKnife VSI, Part Numbers: a) 032500-010: CyberKnife VSI 9.6.x Robotic Radiosurgery System for Japan b) 033000-010: CyberKnife VSI Robotic Radiosurgery System

Reason

There is an unintended drop of a secondary collimator housing that can occur during housing Xchange(TM) workflows. Specifically, during an automated housing Xchange, an interlock may interrupt this operation causing the collimator housing to partially attach.

Code Info

a) 032500-010: CyberKnife VSI 9.6.x Robotic Radiosurgery System for Japan Serial Numbers: C0281, C0307 b) 033000-010: CyberKnife VSI Robotic Radiosurgery System Serial Numbers: C0042, C0064, C0071, C0074, C0079, C0091, C0098, C0102, C0137, C0161, C0180, C0182, C0187, C0200, C0203, C0210, C0216, C0230, C0238, C0246, C0247, C0249, C0252, C0255, C0257, C0264, C0269, C0270, C0273, C0274, C0276, C0277, C0284, C0293, C0294, C0304, C0305, C0322, C0323, C0325, C0326

Distribution

US & Italy, United Kingdom, Switzerland, Belgium, Canada, Germany, France, Japan, Poland, Netherlands, Luxembourg, Australia, Qatar, Latvia, Portugal, India, Hungary, Turkey, United Arab Emirates, Thailand

Quantity

43 units