FDA Enforcement
Class II
Terminated
Brilliance iCT Computed Tomography X-ray system
Recall: Z-1695-2015
·
Reported June 10, 2015
Enforcement
- Recall Number
- Z-1695-2015
- Event ID
- 71231
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Philips Medical Systems (Cleveland) Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 10, 2015
- Initiation Date
- April 13, 2015
- Classification Date
- May 29, 2015
- Termination Date
- October 1, 2015
- Address
- 595 Miner Rd, N/A, Cleveland, OH, 44143-2131, United States
Description
Brilliance iCT Computed Tomography X-ray system
Reason
A calibrated torque wrench malfunctioned at the supplier leading to possible incorrect torque of x-ray tube mounting fasteners.
Code Info
Model# 728306; Serial #'s: 85019 & 85023.
Distribution
Internationally to the countries of China and Norway.
Quantity
2