FDA Enforcement Class II Ongoing

Various Laparoscopy Packs

Recall: Z-1694-2022 · Reported September 14, 2022

Enforcement

Recall Number
Z-1694-2022
Event ID
90574
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
American Contract Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
September 14, 2022
Initiation Date
June 16, 2022
Classification Date
September 8, 2022
Address
2610 Ne Industrial Dr Ste 220, N/A, Kansas City, MO, 64117-2648, United States

Description

Various Laparoscopy Packs

Reason

Product was exposed to multiple sterilization cycles without validation for multiple exposures.

Code Info

Tray Number, Sterilization Lot, Lot Number, Bag Serial Number 1) LSRENDOSCOPH, 2104281, 890211, 14887854 2) LSRENDOSCOPH, 2108251, 784211, 44375070 3) LSRENDOSCOPH, 2110011, 740211, 15140201 4) PLLA88I, 2104162, 895211, 14779054 5) AMGL09X, 2104281, 887211, 64029666 6) IHGB89AB, 2107293, 791211, 64104353 7) UIUA68AG, 2108202, 771211, 64233083 8) FRLC65S, 2104301, 882211, 44123696-005 9) MHLC19AH, 2105181, 863211, 63956744-006 10) CXRO12W, 2107082, 811211, 64110751-006 11) SLLP38F, 2107132, 807211, 64100877-006 12) SLLP38F, 2107132, 807211, 64100879-006 13) SLLP38F, 2107132, 807211, 64100878-006 14) CXLP87AJ, 2107162, 803211, 15011547-004 15) BHLC43J, 2108042, 784211, 44294507-005 16) BHLC43J, 2108042, 784211, 44234508-005 17) MERS50Q, 2108193, 769211, 64206337-006 18) BHRP78C, 2104131, 903211, 64073208-006 19) BHRP78C, 2104131, 903211, 64073209-006 20) BHRP78C, 2106221, 831211, 64091105-006 21) BHRP78C, 2106221, 831211, 64091104-006 22) BHRP78C, 2106233, 831211, 64090598-006 23) BHLP21C, 2107261, 800211, 44226340-005 24) BHLP21C, 2107261, 800211, 44226339-005 25) BHLP21C, 2107261, 800211, 44226341-005 26) BHLP21C, 2107261, 800211, 44226338-005 27) BHLP62E, 2107292, 799211, 44219340-005 28) BHLP62E, 2107292, 799211, 44219343-005 29) BHLP62E, 2107292, 799211, 44219342-005 30) BHLP62E, 2107292, 799211, 44219341-005 31) BHRP78C, 2108041, 789211, 64093909-006 32) BHRP78C, 2108041, 789211, 64092459-006 33) BHRP78C, 2108041, 789211, 64092460-006 34) BHLP62E, 2108242, 771211, 44336911-005 35) SARB30J, 2104291, 884211, 64040187-006 36) MCLC14F, 2106011, 853211, 44153023-005 37) MPRO72D, 2107131, 811211, 64129494-006 38) MCLC14F, 2109031, 760211, 44261569-005 39) AGLC19L, 2108191, 775211, 64186775-006

Distribution

Distribution within the United States to health systems and distributors in AZ, FL, IA, IL, MA, MN, MO, OH, NE, PA, RI, SD, and TX

Quantity

39 units