IntelliVue MX40 Patient Monitor, Product number 865352, Service Numbers All: 453564262571, 453564262591, 453564615311, 453564615331, 453564615351, and 453564615371 Product Usage: Used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network. When the MX40 is connected to the IntelliVue Information Center the IntelliVue Information Center provides the primary patient monitoring and alarming function. The MX40 does not automatically provide local monitoring or alarming when connected to the Information Center.
Enforcement
- Recall Number
- Z-1694-2019
- Event ID
- 82696
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Philips North America, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 12, 2019
- Initiation Date
- March 28, 2019
- Classification Date
- May 31, 2019
- Address
- 3000 Minuteman Rd, Andover, MA, 01810-1032, United States
Description
IntelliVue MX40 Patient Monitor, Product number 865352, Service Numbers All: 453564262571, 453564262591, 453564615311, 453564615331, 453564615351, and 453564615371 Product Usage: Used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network. When the MX40 is connected to the IntelliVue Information Center the IntelliVue Information Center provides the primary patient monitoring and alarming function. The MX40 does not automatically provide local monitoring or alarming when connected to the Information Center.
The MX40 may experience increased power consumption and may have a lack of visual and/or audible notifications while in a Low Battery State .
All revision B.X versions are affected.
Worldwide distribution - US Nationwide and to 71 countries.