FDA Enforcement
Class II
Terminated
Critical Care Ventilator, Catalog Number(s): 8422400: Babylog VN800 - Product Usage: used for treating patients who require temporary or longer term respiratory support for different medical reasons.
Recall: Z-1693-2021
·
Reported June 9, 2021
Enforcement
- Recall Number
- Z-1693-2021
- Event ID
- 87819
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Draeger Medical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 9, 2021
- Initiation Date
- April 16, 2021
- Classification Date
- May 28, 2021
- Termination Date
- February 6, 2024
- Address
- 3135 Quarry Rd, Telford, PA, 18969-1042, United States
Description
Critical Care Ventilator, Catalog Number(s): 8422400: Babylog VN800 - Product Usage: used for treating patients who require temporary or longer term respiratory support for different medical reasons.
Reason
Three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800: 1. Restart of ventilation unit. 2. Incorrect FiO2 high and FiO2 low alarms. 3. SmartCare/PS (SC/PS) suspends weaning
Code Info
All lots/serial numbers manufactured from 1/1/2019-3/31201
Distribution
US Nationwide distribution in the states of PA, TX, MD, FL, TX, WA, MS, NY.
Quantity
12 units