FDA Enforcement
Class II
Terminated
Fresenius Service Replacement 0-Ring. Used in as a service part in 4 sub-assemblies manufactured and/or refurbished Fresenius hemodialysis machines, models 2008K, 2008K2, 2008K@Home and 2008T
Recall: Z-1693-2013
·
Reported July 17, 2013
Enforcement
- Recall Number
- Z-1693-2013
- Event ID
- 65090
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Fresenius Medical Care Holdings, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 17, 2013
- Initiation Date
- May 24, 2013
- Classification Date
- July 11, 2013
- Termination Date
- October 7, 2019
- Address
- 920 Winter St, N/A, Waltham, MA, 02451-1521, United States
Description
Fresenius Service Replacement 0-Ring. Used in as a service part in 4 sub-assemblies manufactured and/or refurbished Fresenius hemodialysis machines, models 2008K, 2008K2, 2008K@Home and 2008T
Reason
Incorrect rubber nitrile O-rings distributed instead of the correct material EPDM. (Ethylene Propylene Dine Monomer (M-class) rubber)
Code Info
Part Number: 640919
Distribution
Nationwide Distribution including the states of AR, CA, FL, GA, KY, LA, MA, MD, NC, NJ, NY, PA, PR, SC, TX and VA.
Quantity
725 units