FDA Enforcement Class II Terminated

Critical Care Ventilator, Catalog Number(s): 8422300: Evita V600 - Product Usage: used for treating patients who require temporary or longer term respiratory support for different medical reasons.

Recall: Z-1690-2021 · Reported June 9, 2021

Enforcement

Recall Number
Z-1690-2021
Event ID
87819
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Draeger Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 9, 2021
Initiation Date
April 16, 2021
Classification Date
May 28, 2021
Termination Date
February 6, 2024
Address
3135 Quarry Rd, Telford, PA, 18969-1042, United States

Description

Critical Care Ventilator, Catalog Number(s): 8422300: Evita V600 - Product Usage: used for treating patients who require temporary or longer term respiratory support for different medical reasons.

Reason

Three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800: 1. Restart of ventilation unit. 2. Incorrect FiO2 high and FiO2 low alarms. 3. SmartCare/PS (SC/PS) suspends weaning

Code Info

All lots/serial numbers manufactured from 1/1/2019-3/31201

Distribution

US Nationwide distribution in the states of PA, TX, MD, FL, TX, WA, MS, NY.

Quantity

34 units