FDA Enforcement Class II Terminated

da Vinci Instrument Control Box, model number ICB3000; Manufactured by Intuitive Surgical, Sunnyvale, CA 94086. Intuitive Surgical endoscopic Instrument Control System is intended to assist in the accurate control of Intuitive Surgical Endoscopic instruments. It is intended for adult and pediatric use.

Recall: Z-1689-2013 · Reported July 17, 2013

Enforcement

Recall Number
Z-1689-2013
Event ID
65590
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Intuitive Surgical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 17, 2013
Initiation Date
June 8, 2013
Classification Date
July 11, 2013
Termination Date
January 22, 2014
Address
1266 Kifer Rd Bldg 100, N/A, Sunnyvale, CA, 94086-5304, United States

Description

da Vinci Instrument Control Box, model number ICB3000; Manufactured by Intuitive Surgical, Sunnyvale, CA 94086. Intuitive Surgical endoscopic Instrument Control System is intended to assist in the accurate control of Intuitive Surgical Endoscopic instruments. It is intended for adult and pediatric use.

Reason

Factory testing on da Vinci Si Surgical Systems may not be in compliance with UL Standards as one of the testing devices was found to be working incorrectly.

Code Info

Part number 372012-96: Serial numbers: 323986 324198.

Distribution

Worldwide Distribution - USA including UT, CA, AZ, CO, FL, HI. MT, NY,TX and WA and one customer in Russia.

Quantity

30 total, all devices