Cyto-Chex BCT - blood specimen collection device - Immunophenotyping Preservative. IVD. Sterile. 2.0ml only. Packaged 6-tube pack REF 213559 and 100-tube box REF 218980. Streck.
Enforcement
- Recall Number
- Z-1684-2017
- Event ID
- 76501
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Streck
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- April 5, 2017
- Initiation Date
- February 23, 2017
- Classification Date
- March 27, 2017
- Address
- 7002 S 109th St, La Vista, NE, 68128-5729, United States
Description
Cyto-Chex BCT - blood specimen collection device - Immunophenotyping Preservative. IVD. Sterile. 2.0ml only. Packaged 6-tube pack REF 213559 and 100-tube box REF 218980. Streck.
Complaints of the 2.0ml tube stopper coming loose during phlebotomy when small diameter tube adapters were not utilized. Such adapters aid with centering small diameter tubes to properly pierce the middle of the stoppers.
UDI Number 10844509000836 and 20844509000833. Lot(s) Expiration Date 50910309 04-01-2017 60760309 03-16-2018 61460309 05-25-2018
Worldwide Distribution - USA (nationwide) and to the countries of : Belgium, Canada, Germany, Israel, Japan, the Republic of Korea, the Netherlands, and the United Kingdom. There was no military or government distribution.
128,692 tubes