FDA Enforcement Class II Ongoing

Product Description/Item: ChemoLock Bag Spike, 10 units/CL-10-10; ChemoLock BAG SPIKE w/ ADDITIVE PORT, DRY SPIKE/CL-12; APPX 1.5ml,ChemoLock BAG SPIKE WITH ADDITIVE PORT/CL-13; ChemoLock Port/CL2100; ChemoLock Port, 5 Units/CL2100-5; ChemoLock Port/CL2100T; Clave Bag Spike w ChemoLock Port/CL2150; 30" (76 cm) Appx 3.3 ml, 20 Drop Admin Set w Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Hanger/CL3011; 9 IN(23cm)APPX 2.7ml,EXT SET,ChemoLock PORT,2 CLAMPS,2 GRADUATED CONN/CL3364; Syringe Transfer Set w Clave, ChemoLock Port/CL-34; 30" (76 cm) Appx 3.6 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, ChemoLock w Red Cap, Bag Hanger/CL3511; 30" (76 cm) Appx 3.6 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, ChemoLock w Red Cap, Bag Hanger/CL3511T; 30" (76cm) Appx 6.3 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, 0.2 Micron Filter, ChemoLock w Red Cap, Hanger/CL3528; 7" (18 cm) Appx .93 ml, Ext Set w/ChemoLock, Y-Connector, Rotating Luer/CL3535T; Oncology Kit w/5" (13 cm) Add-On Set w ChemoLock Additive Port, Vented Cap, ChemoLock, Spinning Spiros w Red Cap/CL3538; Graduated Connector w ChemoLock Port/CL3900; ONCOLOGY KIT, EXT,ChemoLock PORT, GRAD CONN,TRANSFER,ChemoLock,VIAL SPIKE,ChemoLock PORT/CL3927; Oncology Kit w 5" (13 cm) Add-On Set w ChemoLock w Red Cap, Vented Cap, ChemoLock Port w Bag Spike, Spiros w Red Cap/CL3946; Oncology Kit w 17" (43 cm) Bifuse Ext Set w/ChemoLock w Red Cap, ChemoLock Port, ChemoLock Vented Vial Spike, 13mm ChemoLock/CL3947; Oncology Kit w ChemoLock Port Closed Vial Spike, ChemoLock, ChemoLock Port Bag Spike/CL3952; APPX 0.5ml, ChemoLock BAG SPIKE WITH ADDITIVE PORT,ChemoLock Port/CL3955;

Recall: Z-1681-2022 · Reported September 14, 2022

Enforcement

Recall Number
Z-1681-2022
Event ID
90606
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
ICU Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 14, 2022
Initiation Date
June 27, 2022
Classification Date
September 7, 2022
Address
951 Calle Amanecer, N/A, San Clemente, CA, 92673-6212, United States

Description

Product Description/Item: ChemoLock Bag Spike, 10 units/CL-10-10; ChemoLock BAG SPIKE w/ ADDITIVE PORT, DRY SPIKE/CL-12; APPX 1.5ml,ChemoLock BAG SPIKE WITH ADDITIVE PORT/CL-13; ChemoLock Port/CL2100; ChemoLock Port, 5 Units/CL2100-5; ChemoLock Port/CL2100T; Clave Bag Spike w ChemoLock Port/CL2150; 30" (76 cm) Appx 3.3 ml, 20 Drop Admin Set w Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Hanger/CL3011; 9 IN(23cm)APPX 2.7ml,EXT SET,ChemoLock PORT,2 CLAMPS,2 GRADUATED CONN/CL3364; Syringe Transfer Set w Clave, ChemoLock Port/CL-34; 30" (76 cm) Appx 3.6 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, ChemoLock w Red Cap, Bag Hanger/CL3511; 30" (76 cm) Appx 3.6 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, ChemoLock w Red Cap, Bag Hanger/CL3511T; 30" (76cm) Appx 6.3 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, 0.2 Micron Filter, ChemoLock w Red Cap, Hanger/CL3528; 7" (18 cm) Appx .93 ml, Ext Set w/ChemoLock, Y-Connector, Rotating Luer/CL3535T; Oncology Kit w/5" (13 cm) Add-On Set w ChemoLock Additive Port, Vented Cap, ChemoLock, Spinning Spiros w Red Cap/CL3538; Graduated Connector w ChemoLock Port/CL3900; ONCOLOGY KIT, EXT,ChemoLock PORT, GRAD CONN,TRANSFER,ChemoLock,VIAL SPIKE,ChemoLock PORT/CL3927; Oncology Kit w 5" (13 cm) Add-On Set w ChemoLock w Red Cap, Vented Cap, ChemoLock Port w Bag Spike, Spiros w Red Cap/CL3946; Oncology Kit w 17" (43 cm) Bifuse Ext Set w/ChemoLock w Red Cap, ChemoLock Port, ChemoLock Vented Vial Spike, 13mm ChemoLock/CL3947; Oncology Kit w ChemoLock Port Closed Vial Spike, ChemoLock, ChemoLock Port Bag Spike/CL3952; APPX 0.5ml, ChemoLock BAG SPIKE WITH ADDITIVE PORT,ChemoLock Port/CL3955;

Reason

Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic substances. Ports available as standalone connectors, or utilized as access points on vial adaptors, bag spikes, and administration sets.

Code Info

Item/UDI-DI/Lots: CL-10-10/10887709071022/5872839, 5891255; CL-12/10887709056975/5819981, 5856055, 5873186; CL-13/10887709056982/5850811; CL2100/5797867, 5816334, 5829882,5850510; CL2100-5/10887709122960/5754377; CL2100T/10887709099293/5816341, 5842229, 5850517; CL2150/10887709094670/5829900, 5850507; CL3011/10887709057019/5829885, 5842227; CL3364/10887709073408/5841261, 5850805; CL-34/10887709069777/5754437, 5829875; CL3511/10887709057026/5842231, 5861991, 5894477; CL3511T/10887709120010/5829831, 5891263; CL3528/10887709057040/5850500, 5872833, 5883793, 5894481, 5909807; CL3535T/10887709099613/5685350; CL3538/10887709071909/5935851; CL3900/10887709071411/5772697; CL3927/10887709075440/5842211; CL3946/10887709080291/5829895, 5861983; CL3947/10887709080307/5782844; CL3952/10887709081571/5782863,5816379; CL3955/10887709082509/5841209, 5850839; CL3960/10887709083278/5816328, 5842216; CL3963/10887709083827/5842214,5851775; CL4114/10887709091136/5754773, 5782864, 5816356; CL4130/10887709095530/5850515; CL4131/10887709095776/5895940; CL4136/10887709096131/5816325, 5850661; CL4143/10887709097213/5782848, 5873243, 5891266; CL4146-5/10887709097466/5872843, 5883800, 5909812; CL4153/10887709098722/5816348; CL4159/10887709121369/5881696; CL4179/10887709123578/5872687; CL-62/10887709057088/5816320,5825248; CL-70/10887709057095/5842228; CL-70-10/10887709070988/5816343, 5851774; CL-70-10T/10887709099477/5816353; CL-70S/10887709067155/5842225,5850512; CL-70S-5/10887709070995/5816326, 5851776; CL-70T/10887709099507/5816350; CL-72/10887709062440/5829832, 5849740, 5816312; CL-72T/10887709099552/5841181, 5849748; CL-76/10887709122878/5816323, 5842218; CL-80/10887709057125/5804034, 5842200, 5850499; CL-80S/10887709069593/5808063, 5816306, 5829020, 5829023, 5841982, 5845354, 5862839; CL-80S-10/10840619080577/5804045, 5816315; CL-80S-10T/10887709120539/5804051, 5816310, 5829022, 5841994, 5849630, 5850497, 5871336; CL-80S-5/10887709077116/5667054, 5829021, 5829024, 5841980, 5849474, 5849631, 5849632, 5849713, 5849714, 5849715, 5849716, 5849474; CL-80SL/10887709069630/5816317; CL-82/10887709085692/5804065; CLH-12/10887709080413/5842212

Distribution

Worldwide distribution - US Nationwide distribution including in the states of LA, FL, TX, NY, NE, WI, PA, IA, ND, CA, OH, WA, GA, AL, CO, AZ, NC, TN, MI, VA, MA, DE, OR, KY, CT, OK, NJ, MN, IL, VT, AR, RI, ID, IN, UT, NV, MT, MD, KS, SC, NM, ME, MO, SD and the countries of : Saudi Arabia, Australia, United Arab Emirates, Spain, Slovenia, Japan, Canada.

Quantity

454,635