Thermage Face Tip 3.0, REF/UDI: TTNS3,00C3-400/UDI (01)000850608002377; REF TTNS3.00C3-600/UDI (01)000850608002384; REF TTNS3.00C3-900; UDI (01)000850608002391 Used in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

Recall: Z-1681-2020 · Reported April 15, 2020

Enforcement

Recall Number
Z-1681-2020
Event ID
85314
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Solta Medical Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 15, 2020
Initiation Date
December 30, 2019
Classification Date
April 7, 2020
Termination Date
April 29, 2024
Address
11720 N Creek Pkwy N Ste 100, N/A, Bothell, WA, 98011-8244, United States

Description

Thermage Face Tip 3.0, REF/UDI: TTNS3,00C3-400/UDI (01)000850608002377; REF TTNS3.00C3-600/UDI (01)000850608002384; REF TTNS3.00C3-900; UDI (01)000850608002391 Used in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

Reason

Complaints of sparking of the tips of the electrosurgical device used in RF surgical procedures. The use of the affected treatment tips could contribute to potential skin burns.

Code Info

Model Number /Lot Numbers: TTNS3.00C3-400: DOM 120519 011, DOM 120419 014, DOM 120219 008, DOM 112619 006, DOM 112019 010, DOM 111519 004, DOM 100819 005, DOM 100119 005, DOM 082319 003, DOM 081919 007, DOM 081419 004; TTNS3.00C3-600: DOM 120419 018, DOM 112019 009, DOM 110119 010, DOM 092319 005, DOM 082219 015, DOM 082119 006; TTNS3.00C3-900: DOM 111819 005, DOM 092319 006 , DOM 081419 005 Lot number/Number of Units Affected 120219-008/180, 112619-006/225, 112019-010/225, 112019-009/420, 111519-004/220, 110119-010/398, 100819-005/405, 092319-006/23, 092319-005/342, 082319-003/525, 082219-015/504, 082119-006/252, 081419-007/135, 081419-005/126, 081419-004/492 REF/UDI: TTNS3,00C3-400/UDI (01)000850608002377; REF TTNS3.00C3-600/UDI (01)000850608002384; REF TTNS3.00C3-900; UDI (01)000850608002391

Distribution

US - AK, AR, AZ, CA, FL, GA, IL, NY, PA, TX. UT, and WA OUS -Japan, Korea, Malaysia, Singapore, Taiwan, and Thailand.

Quantity

4472 units