FDA Enforcement Class II Terminated

***REF C7122***QTY 1***APEX Arthroscopy Tubing Set. Intended for use for joint irrigation and dissention during arthroscopic procedures.

Recall: Z-1681-2013 · Reported July 17, 2013

Enforcement

Recall Number
Z-1681-2013
Event ID
65320
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Linvatec Corp. dba ConMed Linvatec
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
July 17, 2013
Initiation Date
May 30, 2013
Classification Date
July 10, 2013
Termination Date
September 4, 2013
Address
11311 Concept Blvd, N/A, Largo, FL, 33773-4908, United States

Description

***REF C7122***QTY 1***APEX Arthroscopy Tubing Set. Intended for use for joint irrigation and dissention during arthroscopic procedures.

Reason

Conmed Linvatec received complaints that their products C7120 APEX Arthroscopy Tubing Set, One Connection and C7122 APEX Arthroscopy Tubing Set, Two Connection, contained a scratch/grooved area on the inside diameter of the pressure fitting that may cause the APEX Irrigation (pump) Console to sense a pressure drop and increase fluid flow to the surgical site.

Code Info

C7122 Lots #'s: 1301224, 1301234, 1301244, 1301304, 1301314, 1302114, 1302154, 1302274, 1302284, 1303144, 1303254, 1304044, 1304234, 1304244, and 1305024

Distribution

Worldwide Distribution-USA (nationwide) including the states of AR, CA, FL, GA, IL, IN, KS, MO, NC, NJ, NM, NY, OH, OK, PA, TN, TX, UT, VA, and WI, and the countries of Austria; Belgium, Canada, China, Czech Republic, Finland, France, Great Britain, Italy, and Japan.

Quantity

4,691 units, total of each