FDA Enforcement Class II Ongoing

Azurion 5 M20; System Code: (1)722228, (2)722232, (3) 722281 (OUS only);

Recall: Z-1680-2026 · Reported April 8, 2026

Enforcement

Recall Number
Z-1680-2026
Event ID
98434
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 8, 2026
Initiation Date
March 3, 2026
Classification Date
March 27, 2026
Address
Veenpluis 6, Best, N/A, Netherlands

Description

Azurion 5 M20; System Code: (1)722228, (2)722232, (3) 722281 (OUS only);

Reason

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

Code Info

(1) System Code: 722228; UDI: 00884838099234; System Serial Number: 67, 524, 49, 678, 679, 10, 36, 64, 110, 403, 328, 481, 373, 149, 213, 45, 537, 615, 159, 18, 231, 180, 61, 204, 273, 538, 76, 502, 150, 132, 399, 533, 211, 17, 47, 38, 87, 617, 618, 619, 128, 522, 32, 600, 23, 25, 62, 177, 181, 48, 189, 153, 79, 80, 157, 24, 187, 26, 642, 431, 29, 113, 122, 127, 6, 123, 124, 185, 186, 198, 92, 215, 573, 91, 529, 523, 205, 20, 37, 659, 145, 3, 567, 539, 645, 156, 308, 333, 335, 570, 560, 44, 569, 418, 525, 280, 616, 387, 568, 471, 428, 465, 556, 629, 102, 103, 176, 178, 109, 16, 299, 234, 566, 508, 571, 463, 71, 493, 1, 434, 647, 317, 562, 230, 293, 552, 457, 458, 369, 561, 517, 203, 554, 260, 100, 101, 107, 108, 640, 99, 318, 21, 22, 676, 677, 154, 531, 75, 602, 78, 574, 158, 461, 370, 613, 393, 564, 482, 194, 660, 673, 60, 464, 4, 462, 138, 223, 312, 319, 98, 623, 475, 125, 218, 513, 442, 435, 9, 95, 557, 174, 548, 486, 487, 84, 454, 346, 396, 39, 339, 172, 191, 327, 52, 575, 196, 406, 118, 415, 436, 114, 452, 53, 216, 761, 179, 398, 31, 88, 331, 297, 392, 374, 503, 412, 385, 343, 427, 488, 389, 349, 536, 214, 232, 500, 713, 622, 727, 469, 588, 365, 443, 510, 757, 758, 692, 747, 705, 414, 712, 728, 621, 710, 715, 726, 717, 90, 706, 417, 709, 498, 729, 130, 707, 714, 718, 755, 711, 716, 750, 708, 744, 704, 284, 663, 725, 565, 745, 73, 474, 724, 735, 703, 201, 407, 397, 104, 202, 690, 46, 171, 34, 70, 680, 657, 229, 259, 332, 643, 501, 286, 306, 610, 117, 646, 13, 395, 429, 12, 56, 654, 460, 601, 579, 57, 86, 665, 628, 257, 607, 608, 430, 360, 650, 651, 590, 591, 595, 352, 329, 197, 366, 345, 627, 302, 166, 664, 432, 361, 305, 542, 313, 620, 220, 325, 545, 405, 514, 433, 341, 342, 184, 424, 425, 426, 497, 630, 258, 183, 195, 314, 555, 626, 282, 402, 490, 439, 421, 440, 446, 338, 472, 662, 689, 633, 320, 281, 480, 631, 364, 581, 182, 448, 283, 291, 473, 321, 476, 681, 409, 535, 376, 606, 519, 655, 384, 534, 362, 520, 489, 592, 303, 413, 496, 580, 668, 106, 63, 685, 691, 701, 722, 723, 730, 732, 733, 765, 35, 512, 404, 74, 437, 614, 546, 14, 336, 200, 344, 356, 357, 394, 7, 504, 190, 377, 383, 193, 94, 89, 93, 309, 310, 494, 444, 199, 667, 391, 408, 532, 698, 83, 143, 518, 85, 483, 449, 337, 438, 453, 311, 324, 639, 353, 459, 5, 43, 477, 330, 386, 288, 167, 549, 390, 368, 192, 66, 115, 116, 334, 467, 307, 304, 350, 212, 340, 348, 506, 516, 515, 292, 33, 173, 543, 217, 219, 160, 702, 634, 82, 135, 133, 208, 478, 162, 165, 635, 597, 378, 466, 163, 168, 479, 164, 81, 228, 401, 688, 97, 719, 139, 540, 121, 147, 354, 141, 734, 285, 720, 65, 140, 598, 294, 146, 175, 170, 637, 295, 290, 134, 136, 316, 289, 207, 686, 492, 416, 298, 30, 59, 419, 11, 119, 28, 144, 68, 55, 27, 233, 611; (2) System Code: 722232; UDI: 00884838116757; System Serial Number: 5, 33, 34, 106, 49, 138, 91, 2, 95, 96, 36, 23, 108, 42, 48, 46, 76, 6, 60, 61, 10, 43, 97, 8, 113, 118, 119, 139, 22, 50, 75, 63, 130, 132, 89, 86, 69, 3, 67, 39, 4, 44, 122, 123, 71, 72, 110, 93, 1, 68, 11, 77, 20, 7, 57, 58, 260, 115, 14, 82, 70, 112, 81, 131, 277, 128, 31, 37, 120, 121, 35, 293, 28, 30, 54, 102, 19, 134, 137, 29, 56, 51, 32, 105, 16, 117, 73, 100, 13; (3) System Code: 722281; UDI: 00884838110564; System Serial Number: 125, 176, 233, 296, 74, 284, 55, 81, 37, 21, 135, 278, 390, 179, 181, 182, 282, 286, 357, 358, 359, 109, 136, 93, 153, 247, 27, 28, 299, 76, 140, 60, 341, 310, 192, 275, 127, 108, 254, 255, 24, 20, 129, 343, 295, 215, 66, 98, 48, 235, 378, 145, 336, 115, 348, 120, 17, 12, 65, 217, 339, 351, 95, 364, 236, 308, 44, 45, 319, 164, 124, 159, 317, 318, 64, 189, 71, 38, 387, 228, 272, 54, 19, 26, 105, 344, 259, 260, 261, 316, 80, 265, 70, 190, 225, 186, 180, 281, 202, 349, 149, 117, 388, 52, 175, 187, 270, 11, 219, 371, 269, 157, 290, 102, 244, 94, 110, 23, 9, 208, 209, 168, 204, 114, 253, 35, 155, 199, 201, 161, 126, 211, 287, 207, 85, 321, 227, 276, 226, 169, 156, 258, 170, 415, 294, 101, 229, 146, 14, 88, 99, 188, 89, 273, 340, 68, 151, 141, 221, 33, 307, 350, 289, 246, 83, 10, 177, 39, 40, 43, 326, 331, 57, 212, 304, 122, 178, 367, 323, 69, 191, 50, 13, 61, 194, 200, 250, 251, 34, 132, 404, 325, 205, 82, 222, 138, 174, 311, 312, 121, 300, 90, 96, 231, 123, 298, 160, 335, 75, 91, 292, 369, 223, 18, 86, 87, 49, 385, 166, 72, 274, 47, 242, 324, 144, 355, 158, 234, 345, 22, 224, 197, 198, 206, 329, 118, 249, 58, 133, 263, 193, 280, 137, 303, 302, 103, 353, 196, 139, 173, 143, 29, 239, 15, 163, 172, 111, 252, 230, 42, 213, 220, 346, 347, 362, 185, 30, 31, 25, 106, 142, 36, 84, 16, 257, 241, 243, 262, 165, 147, 248, 288;

Distribution

Domestic: Nationwide Distribution; International: Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, FrenchPolynesia, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Lao, Laos, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkiye, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen;

Quantity

920 units (168 US, 752 OUS)