FDA Enforcement Class II Terminated

115 V Blanketrol II, Model 222S Hyper-Hypothermia System . Used to lower or to raise a patients temperature and /or maintain a desired patient temperature through conductive heat transfer. Part number 86165.

Recall: Z-1679-2014 · Reported June 11, 2014

Enforcement

Recall Number
Z-1679-2014
Event ID
68356
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Cincinnati Sub-Zero Products Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
June 11, 2014
Initiation Date
January 8, 2014
Classification Date
June 2, 2014
Termination Date
June 6, 2014
Address
12011 Mosteller Rd, Cincinnati, OH, 45241-1528, United States

Description

115 V Blanketrol II, Model 222S Hyper-Hypothermia System . Used to lower or to raise a patients temperature and /or maintain a desired patient temperature through conductive heat transfer. Part number 86165.

Reason

The incorrect voltage component was placed into the device during manufacturing. This may make the device unable to maintain patient temperature during cooling mode. The device will continue to indicate that the unit is cooling because the compressor will continue to run, however water in the reservoir will begin to return to ambient temperature. The actual water temperature will be displayed but

Code Info

Model Number 222S; Serial #s: 133-2-00697, 133-2-00698; 133-2-00699; 133-2-00700; 133-2-00701; 133-2-00702; and 133-2-00703:.Part # 86165

Distribution

The product was distributed to one customer in MI, and internationally (Taiwan)

Quantity

7 units