FDA Enforcement Class II Ongoing

VITROS XT 7600 Integrated System with Version 3.7.2 and below Product Code: 6844461

Recall: Z-1676-2022 · Reported September 14, 2022

Enforcement

Recall Number
Z-1676-2022
Event ID
90681
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Ortho-Clinical Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 14, 2022
Initiation Date
July 19, 2022
Classification Date
September 2, 2022
Address
100 Indigo Creek Dr, N/A, Rochester, NY, 14626-5101, United States

Description

VITROS XT 7600 Integrated System with Version 3.7.2 and below Product Code: 6844461

Reason

Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction with the Digital Imaging Reflectometer. Potential exists for either a positive or negative shift in results reported by the VITROS System.

Code Info

UDI-DI: 10758750031610 Systems start with 7600XXXX where XXXX is unique for each analyzer

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda DV04, Brazil, Canada L3R 4G5 , Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom.

Quantity

1148 units