VITROS XT 7600 Integrated System with Version 3.7.2 and below Product Code: 6844461
Enforcement
- Recall Number
- Z-1676-2022
- Event ID
- 90681
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Ortho-Clinical Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 14, 2022
- Initiation Date
- July 19, 2022
- Classification Date
- September 2, 2022
- Address
- 100 Indigo Creek Dr, N/A, Rochester, NY, 14626-5101, United States
Description
VITROS XT 7600 Integrated System with Version 3.7.2 and below Product Code: 6844461
Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction with the Digital Imaging Reflectometer. Potential exists for either a positive or negative shift in results reported by the VITROS System.
UDI-DI: 10758750031610 Systems start with 7600XXXX where XXXX is unique for each analyzer
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda DV04, Brazil, Canada L3R 4G5 , Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom.
1148 units