Straight Inflow/Outflow Patient Line Kit. Model Number: 902-00037. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Straight Inflow/Outflow Patient Line Kit (P/N 902-00037) consists of 4.4 Liter Reservoir (P/N 902-00034P), Heat Exchanger set (P/N 902-00006P) and Straight Inflow/Outflow Patient Line (P/N 902-00039P). The three components are packaged separately in individual pouches and boxes and placed in the same case for shipment. All three components are connected prior to use. The female quick connector that may be cracked is located on the Heat Exchanger set (P/N 902-00006P).
Enforcement
- Recall Number
- Z-1675-2025
- Event ID
- 96617
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Belmont Instrument LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 7, 2025
- Initiation Date
- March 31, 2025
- Classification Date
- April 25, 2025
- Address
- 780 Boston Rd, Billerica, MA, 01821-5925, United States
Description
Straight Inflow/Outflow Patient Line Kit. Model Number: 902-00037. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Straight Inflow/Outflow Patient Line Kit (P/N 902-00037) consists of 4.4 Liter Reservoir (P/N 902-00034P), Heat Exchanger set (P/N 902-00006P) and Straight Inflow/Outflow Patient Line (P/N 902-00039P). The three components are packaged separately in individual pouches and boxes and placed in the same case for shipment. All three components are connected prior to use. The female quick connector that may be cracked is located on the Heat Exchanger set (P/N 902-00006P).
Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser.
Model Number: 902-00037. UDI-DI: 00896128002213. Lot Numbers: 20241012, 20241107
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, CT, DC, IL, IN, KS, KY, MA, ME, MN, NC, ND, NJ, NM, OH, OR, PA, SC, TN, TX, VA, VT, WA, WI, WV and the countries of Canada, Hungary, Korea (the Republic of), Latvia, Malaysia.
208 sets