FDA Enforcement Class II Terminated

Integra Licox Test Kit REF BC10R Rx Only An accessory provided for use with the LCXO2 and AC31 PtO2 Monitors. It provides users with a secondary means to verify the monitor's PtO2 state-of-calibration (but does not provide calibration) when needed by providing a specific electrical signal to the monitor.

Recall: Z-1675-2014 · Reported June 11, 2014

Enforcement

Recall Number
Z-1675-2014
Event ID
68159
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Integra LifeSciences Corp.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 11, 2014
Initiation Date
April 16, 2014
Classification Date
May 30, 2014
Termination Date
October 24, 2014
Address
311 Enterprise Dr, Plainsboro, NJ, 08536-3344, United States

Description

Integra Licox Test Kit REF BC10R Rx Only An accessory provided for use with the LCXO2 and AC31 PtO2 Monitors. It provides users with a secondary means to verify the monitor's PtO2 state-of-calibration (but does not provide calibration) when needed by providing a specific electrical signal to the monitor.

Reason

Integra LifeSciences has identified through a small number of complaints that some lot numbers of Licox Test Kits (Catalog Number BC10R) may provide erroneous "calibration test failure" message when they are used with the Licox PtO2 Monitor (LCSO2, or may provide an erroneous out of range result with the Licox CMP Brain Tissue Oxygen and Temperature Monitor (AC31).

Code Info

Catalog Number BC10R Lot Nos: 051113, 100913, 140513, 161013, 25013

Distribution

Worldwide Distribution - USA including Illinois, Florida, Maryland, Massachusetts, North Carolina, Pennsylvania, South Dakota and Internationally to Australia, Austria, Belgium, Chile, Finland, Germany, Great Britain, Greece, Israel, Italy and Russia.

Quantity

49 test probes