FDA Enforcement Class II Terminated

Beckman Coulter MicroScan PROMPT Inoculation System-D, Catalog Number B1026-10D, UDI: 15099590659561 - Product Usage: The Prompt Inoculation System-D is used to standardize inocula for microdilution antimicrobial susceptibility test.

Recall: Z-1674-2020 · Reported April 15, 2020

Enforcement

Recall Number
Z-1674-2020
Event ID
85137
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Beckman Coulter Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
April 15, 2020
Initiation Date
December 13, 2019
Classification Date
April 6, 2020
Termination Date
April 30, 2021
Address
250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States

Description

Beckman Coulter MicroScan PROMPT Inoculation System-D, Catalog Number B1026-10D, UDI: 15099590659561 - Product Usage: The Prompt Inoculation System-D is used to standardize inocula for microdilution antimicrobial susceptibility test.

Reason

Due to customer reports it has been determined that the inoculation system contains the incorrect prompt wands. Use of the incorrect wands could result in false -susceptible results, false-resistant results, possible misidentifications, or possible out-of-range Quality Control results.

Code Info

Lot number 20201103T, Expiration Date: 11/3/2020

Distribution

Worldwide distribution - US Nationwide distribution in the states of AL, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, MD, MI, MS, NE, NY, NC, ND, OH, OK, PA, SD, TN, TX, UT, WV, WI, WY and countries of Canada, China, Japan, Paraguay.

Quantity

3,472 units