LabPro Data Management System, UDI: 15099590676872 - Product Usage: The LabPro Data Management System is a Microsoft Windows based software program and is intended to manage both microbial identification (ID) and antimicrobial agent susceptibility testing (AST) data generated from MicroScan instruments or manually entered microbiology test results, for use by trained laboratory personnel.
Enforcement
- Recall Number
- Z-1673-2020
- Event ID
- 85048
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Beckman Coulter Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- April 15, 2020
- Initiation Date
- January 17, 2020
- Classification Date
- April 6, 2020
- Termination Date
- April 27, 2021
- Address
- 250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States
Description
LabPro Data Management System, UDI: 15099590676872 - Product Usage: The LabPro Data Management System is a Microsoft Windows based software program and is intended to manage both microbial identification (ID) and antimicrobial agent susceptibility testing (AST) data generated from MicroScan instruments or manually entered microbiology test results, for use by trained laboratory personnel.
A security-only update was released via remote diagnostics on January 14, 2020 to customers with Windows 7 operation system. Shortly after the releases some customers reported that Data Management System computers would not restart. If the computer is unable to restart, then the system cannot download data generated from MicroScan instruments or manually entered microbiology test results, resulting in a potential of delayed results.
Computers with Remote Diagnostics (Axeda) enabled and Windows 7 operating system.
Worldwide distribution - US Nationwide distribution in the states of Wisconsin, New York, Delaware, Indiana, Ohio, Tennessee, Florida, Georgia, Illinois, South Carolina, California, New Jersey, Alabama, Massachusetts, Michigan, Texas, West Virginia, Wyoming, Missouri, New Mexico, North Carolina, Minnesota, New Hampshire, Oklahoma, Washington, Pennsylvania, Idaho, Mississippi, Maryland, Virginia, Arizona, Utah, Maine, Michigan, Colorado, Iowa, Louisiana, Kentucky, South Dakota, Nevada, North Dakota, Kentucky and countries of Brazil, Germany.
177 systems