ARCHITECT C System Mixer - Product Usage: The ARCHITECT Clinical Chemistry System is designed to perform automated chemistry tests, utilizing photometry and potentiometric technology. The mixer is a component and it is used on all ARCHITECT Clinical Chemistry systems to mix the sample with the reagent inside of the cuvettes.
Enforcement
- Recall Number
- Z-1671-2020
- Event ID
- 84995
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Abbott Laboratories, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 15, 2020
- Initiation Date
- May 1, 2019
- Classification Date
- April 6, 2020
- Termination Date
- July 2, 2021
- Address
- 1921 Hurd Dr, N/A, Irving, TX, 75038-4313, United States
Description
ARCHITECT C System Mixer - Product Usage: The ARCHITECT Clinical Chemistry System is designed to perform automated chemistry tests, utilizing photometry and potentiometric technology. The mixer is a component and it is used on all ARCHITECT Clinical Chemistry systems to mix the sample with the reagent inside of the cuvettes.
Mixer blade may separate from mixer due to the screw and nut failure and detachment. As a result of the mixer blade missing, the sample and reagent are inadequately mixed.
LN 09D59-01, LN 09D59-02, LN 09D59-03
Worldwide distribution US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, ME, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, Puerto Rico and St. Croix US Virgin Islands. Countries of AF, AL, AO, AR, AM, AU, AT, AZ, BS, BD, BV, BE, BA, BW, BR, BN, BG, BF, CA, CL, CN, CO, CR, HR, CY, CZ, CD, DK, DO, EC, EG, SV, EE, ET, FI, FR, GE, DE, GR, GT, HN, HK, HU, IN, ID, IQ, IE, IL, IT, JM, JO, KZ, KE, KW, KG, LV, LB, LY, LT, LU, MK, MW, MY, MV, ML, MR, MX, MD, MA, MM, NA, NL, NZ, NG, NO, OM, PK, PA, PE, PH, PL, PT, PY, QA, RO, RU, RW, SA, SN, SG, SK, SI, ZA, ES, SE, CH, TW, TJ, TZ, TH, TT, TN, TC, TR, AE, UG, UK, UA, UY, VE, VN, YE, ZM, ZW, LC, MS, ME, VC, LK, RS, KY, CW, GZA, JRH, CI, KR.
09D59-01, 1010 products, 09D59-02, 25156 products, 09D59-03 13976 products