FDA Enforcement Class II Ongoing

Smith&Nephew ENGAGE Cementless Partial Knee System, Porous Femoral components, Part Numbers: a) REF 1-10003-100 (SIZE 1 Right Medial); b) REF 1-10003-110 (SIZE 1 Left Medial); c) REF 1-10003-200 (SIZE 2-Right Medial); d) REF 1-10003-210 (SIZE 2-Left Medial); e) REF 1-10003-300 (SIZE 3- Right Medial); f) REF 1-10003-310 (SIZE 3- Left Medial); g) REF 1-10003-400 (SIZE 4- Right Medial); h) REF 1-10003-410 (SIZE 4- Left Medial); i) REF 1-10003-500 (SIZE 5- Right Medial); j) REF 1-10003-510 (SIZE 5- Left Medial); k) REF 1-10003-600 (SIZE 6- Right Medial); l) REF 1-10003-610 (SIZE 6- Left Medial); m) REF 1-10003-700 (SIZE 7- Right Medial); n) REF 1-10003-710 (SIZE 7- Left Medial); o) REF 1-10003-800, (SIZE 8 - Right Medial); p) REF 1-10003-810, (SIZE 8 - Left Medial); Unicondylar knee prothesis

Recall: Z-1670-2023 · Reported June 7, 2023

Enforcement

Recall Number
Z-1670-2023
Event ID
92286
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Smith & Nephew, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
June 7, 2023
Initiation Date
May 2, 2023
Classification Date
May 31, 2023
Address
1450 E Brooks Rd, N/A, Memphis, TN, 38116-1804, United States

Description

Smith&Nephew ENGAGE Cementless Partial Knee System, Porous Femoral components, Part Numbers: a) REF 1-10003-100 (SIZE 1 Right Medial); b) REF 1-10003-110 (SIZE 1 Left Medial); c) REF 1-10003-200 (SIZE 2-Right Medial); d) REF 1-10003-210 (SIZE 2-Left Medial); e) REF 1-10003-300 (SIZE 3- Right Medial); f) REF 1-10003-310 (SIZE 3- Left Medial); g) REF 1-10003-400 (SIZE 4- Right Medial); h) REF 1-10003-410 (SIZE 4- Left Medial); i) REF 1-10003-500 (SIZE 5- Right Medial); j) REF 1-10003-510 (SIZE 5- Left Medial); k) REF 1-10003-600 (SIZE 6- Right Medial); l) REF 1-10003-610 (SIZE 6- Left Medial); m) REF 1-10003-700 (SIZE 7- Right Medial); n) REF 1-10003-710 (SIZE 7- Left Medial); o) REF 1-10003-800, (SIZE 8 - Right Medial); p) REF 1-10003-810, (SIZE 8 - Left Medial); Unicondylar knee prothesis

Reason

Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.

Code Info

a) REF 1-10003-100, UDI/DI 00885556872925; b) REF 1-10003-110, UDI/DI 00885556872932; c) REF 1-10003-200, UDI/DI 00885556872949; d) REF 1-10003-210, UDI/DI 00885556872956; e) REF 1-10003-300, UDI/DI 00885556872963; f) REF 1-10003-310, UDI/DI 00885556872970; g) REF 1-10003-400, UDI/DI 00885556872987; h) REF 1-10003-410, UDI/DI 00885556872994; i) REF 1-10003-500, UDI/DI 00885556873007; j) REF 1-10003-510, UDI/DI 00885556873014; k) REF 1-10003-600, UDI/DI 00885556873021; l) REF 1-10003-610, UDI/DI 00885556873038; m) REF 1-10003-700, UDI/DI 00885556873045; n) REF 1-10003-710, UDI/DI 00885556873052; o) REF 1-10003-800, UDI/DI 00885556873069; p) REF 1-10003-810, UDI/DI 00885556873076; ALL BATCHES

Distribution

US Nationwide distribution.

Quantity