SIEMENS Uroskop Omnia Max ; a solid state detector fluoroscopic X-Ray system, primarily for urological applications (functional x-ray diagnostic, endourology and minimal invasive urology/surgery). The system,, which includes a radiologic/urologic treatment table, may be used for urological, gastroenterological and gynecological treatment, planning and diagnostic procedures including but not limited to: Querying and retrieving patient history information and/or previous diagnosis and images from other modalities, including X-ray examinations of the urogenital area, Ultrasound examinations, Endourological interventions, Percutaneous interventions, Laparoscopy, Application of fistula, Simple procedures, Extracorporeal shock wave lithotripsy, Uroflow/urodynamics, Pediatric radiological and therapeutic applications.

Recall: Z-1670-2015 · Reported June 3, 2015

Enforcement

Recall Number
Z-1670-2015
Event ID
71115
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 3, 2015
Initiation Date
April 13, 2015
Classification Date
May 23, 2015
Termination Date
March 20, 2017
Address
51 Valley Stream Pkwy, Malvern, PA, 19355-1406, United States

Description

SIEMENS Uroskop Omnia Max ; a solid state detector fluoroscopic X-Ray system, primarily for urological applications (functional x-ray diagnostic, endourology and minimal invasive urology/surgery). The system,, which includes a radiologic/urologic treatment table, may be used for urological, gastroenterological and gynecological treatment, planning and diagnostic procedures including but not limited to: Querying and retrieving patient history information and/or previous diagnosis and images from other modalities, including X-ray examinations of the urogenital area, Ultrasound examinations, Endourological interventions, Percutaneous interventions, Laparoscopy, Application of fistula, Simple procedures, Extracorporeal shock wave lithotripsy, Uroflow/urodynamics, Pediatric radiological and therapeutic applications.

Reason

Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.

Code Info

Model Number of device 10762473 with serial numbers: 4014 4056 4015 4063 4022 4036 4043 4024 4059 4020 4032 4053 4061 4060 4051 4025 4037 4054 4042 4033 4055 4052 4047 4049 4038 4062

Distribution

Nationwide Distribution.

Quantity

133 total