FDA Enforcement Class II Ongoing

Intelli-C, Left side suspended Tabletop, REF: 03400010

Recall: Z-1669-2023 · Reported June 7, 2023

Enforcement

Recall Number
Z-1669-2023
Event ID
92230
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
NRT X-RAY A/S
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
June 7, 2023
Initiation Date
April 27, 2023
Classification Date
May 31, 2023
Address
Birkegardsvej 16, N/A, Hasselager, N/A, N/A, Denmark

Description

Intelli-C, Left side suspended Tabletop, REF: 03400010

Reason

There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.

Code Info

UDI-DI:05713464000121/Serial Numbers: 647009, 651002, 651006, 666009, 682003

Distribution

US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, LA, NV, NY.

Quantity

13 units