FDA Enforcement Class II Ongoing

Intelli-C, Right side suspended Tabletop, REF: 03400000

Recall: Z-1668-2023 · Reported June 7, 2023

Enforcement

Recall Number
Z-1668-2023
Event ID
92230
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
NRT X-RAY A/S
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
June 7, 2023
Initiation Date
April 27, 2023
Classification Date
May 31, 2023
Address
Birkegardsvej 16, N/A, Hasselager, N/A, N/A, Denmark

Description

Intelli-C, Right side suspended Tabletop, REF: 03400000

Reason

There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.

Code Info

UDI-DI: 05713464000527/Serial Numbers: 627004, 627007, 639001, 639002,639003, 639004, 639007, 639010, 647001, 647004, 647005, 648001, 651007, 658002, 658003, 658006, 658010, 666008, 675001, 675004, 682006, 682010

Distribution

US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, LA, NV, NY.

Quantity

32 units