FDA Enforcement Class I Ongoing

StealthStation Cranial Software, Model: 9735585, used with StealthStation S7/i7 systems

Recall: Z-1666-2023 · Reported June 14, 2023

Enforcement

Recall Number
Z-1666-2023
Event ID
92269
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Medtronic Navigation, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 14, 2023
Initiation Date
April 11, 2023
Classification Date
June 2, 2023
Address
200 Medtronic Drive, Lafayette, CO, 80026, United States

Description

StealthStation Cranial Software, Model: 9735585, used with StealthStation S7/i7 systems

Reason

In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain stimulation, where distance to/past target text no longer synchronized with navigation information, inaccurate value displayed; may result in prolonged/additional procedure, tissue injury

Code Info

Model/UDI-DI/Software Versions: 9735585/00763000253165, 00763000306762 (inside Kits: 00763000299927, 00763000299972), 00763000517267(inside Kits: 00763000517274, 00763000517335)/v3.1.1, 3.1.2, and 3.1.3

Distribution

US Distribution: PR. OUS (foreign) distribution: ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BOSNIA and HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CANARY ISLANDS, CHILE, Colombia COLOMBIA, C¿TE D'IVOIRE, CROATIA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE, HUNGARY, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KENYA, LATVIA, MEXICO, MOLDAVIA, NETHERLANDS, PAKISTAN, POLAND, PORTUGAL, REPUBLIC of KOREA, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, THAILAND, TURKEY, UNITED ARAB EMIRATES

Quantity

556