FDA Enforcement Class II Terminated

Baxter Sigma Spectrum Infusion Pump with Master Drug Library (Version 6) 35700BAXRP Product Usage: The Baxter Sigma Spectrum Infusion Pump with master Drug Library is intended to be used for the controlled administration of fluids. These fluids may includes pharmaceutical drugs, blood, blood products and mixture if required patient therapy. The intended routes of administration consist of the following clinically accepted routes: intravenous, arterial, subcutaneous, intrathecal, epidural or irrigation of fluids. Spectrum infusion pump is intended to be used in conjunction with legally marketed intravenous administration sets and medications provided by user. The Spectrum and Spectrum with MAster Drug Library is intended to reduce operator interaction through automated programming thereby helping to reduce errors associated with complex device programming.

Recall: Z-1666-2018 · Reported May 16, 2018

Enforcement

Recall Number
Z-1666-2018
Event ID
79815
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Baxter Healthcare Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 16, 2018
Initiation Date
April 3, 2018
Classification Date
May 4, 2018
Termination Date
February 14, 2020
Address
711 Park Ave, N/A, Medina, NY, 14103-1036, United States

Description

Baxter Sigma Spectrum Infusion Pump with Master Drug Library (Version 6) 35700BAXRP Product Usage: The Baxter Sigma Spectrum Infusion Pump with master Drug Library is intended to be used for the controlled administration of fluids. These fluids may includes pharmaceutical drugs, blood, blood products and mixture if required patient therapy. The intended routes of administration consist of the following clinically accepted routes: intravenous, arterial, subcutaneous, intrathecal, epidural or irrigation of fluids. Spectrum infusion pump is intended to be used in conjunction with legally marketed intravenous administration sets and medications provided by user. The Spectrum and Spectrum with MAster Drug Library is intended to reduce operator interaction through automated programming thereby helping to reduce errors associated with complex device programming.

Reason

Service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at 400ml/hr and 800ml/hr and verification of upstream occlusion detection at 800ml/hr. Improper detection of air-in-line or upstream occlusion may result in serious adverse health consequences.

Code Info

Serial numbers 792177 795979 802612 805938 807209 813365 813367 871387 922364 935968 962932 963087

Distribution

US Nationwide & Puerto Rico

Quantity

444 units in total