FDA Enforcement
Class II
Ongoing
Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
Recall: Z-1663-2025
·
Reported April 30, 2025
Enforcement
- Recall Number
- Z-1663-2025
- Event ID
- 96581
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- IHB OPERATIONS B.V.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- April 30, 2025
- Initiation Date
- March 27, 2025
- Classification Date
- April 23, 2025
- Address
- Floresstraat 52, N/A, Zwolle, N/A, N/A, Netherlands
Description
Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
Reason
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
Code Info
UDI/DI 10841736116071, All Serial Numbers
Distribution
US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.
Quantity
112 units