Teleflex HORIZON Microclip Titanium Ligating Clips w/tape, REF 005200; for use in general surgical procedures that require vessel or tissue ligation
Enforcement
- Recall Number
- Z-1663-2023
- Event ID
- 92212
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- TELEFLEX LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 7, 2023
- Initiation Date
- April 28, 2023
- Classification Date
- May 30, 2023
- Address
- 3015 Carrington Mill Blvd, N/A, Morrisville, NC, 27560-5437, United States
Description
Teleflex HORIZON Microclip Titanium Ligating Clips w/tape, REF 005200; for use in general surgical procedures that require vessel or tissue ligation
Teleflex LLC is initiating this voluntary field corrective action because the clip cartridges for the affected product is missing a key feature that restricts the application of the clip to the applier. With this step feature missing on the cartridge, the clip is mounted in a more variable position, allowing the clips to load farther back in the clip applier jaw grooves. This may potentially result in inappropriate or inadequate clip application to vascular structures.
UDI/DI 24026704696394, Lot Numbers: 73A2200857, 73A2300018, 73A2300176, 73A2300326, 73A2300520, 73A2300778, 73D2200216, 73D2200559, 73E2200634, 73E2200874, 73E2201099, 73F2200193, 73F2200200, 73F2200482, 73F2200711, 73G2200168, 73G2200345, 73G2200560, 73G2200754, 73H2200115, 73H2200294, 73H2200433, 73H2200453, 73J2200747, 73J2200763, 73K2200014, 73K2200095, 73K2200101, 73K2200299, 73K2200566, 73K2200713, 73L2200064, 73L2200066
worldwide
146,606 units