Maquet Getinge- BO-T 41301 BUMC ECC PACK Material: -701065209R01
Recall: Z-1662-2019
·
Reported June 5, 2019
Enforcement
- Recall Number
- Z-1662-2019
- Event ID
- 82703
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Maquet Cardiovascular, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 5, 2019
- Initiation Date
- April 18, 2019
- Classification Date
- May 30, 2019
- Termination Date
- July 28, 2025
- Address
- 45 Barbour Pond Dr, N/A, Wayne, NJ, 07470-2094, United States
Description
Maquet Getinge- BO-T 41301 BUMC ECC PACK Material: -701065209R01
Reason
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
Code Info
Batch Numbers: 3000060743 3000066277 3000069078 3000073171 3000074176 3000076468 3000080999 3000081905 3000085277
Distribution
Nationwide
Quantity
N/A